VENTImotion 2 VENTImotion 2 with VENTIclick – Actual values: recalculated after every breath (no averaging) – Mean values: calculated across all breaths since the device was started up – Leakage: calculated continuously, updated after every breath – AirTrap statistics: calculated across all breaths since the device was started – Volume compensation: in the “slow” stage, the device checks after every 8 breaths whether the target volume has been reached and changes pressure by 0.5 hPa. If the pressure reaches a corridor around the Filters and smoothing techniques target volume, the device switches to precise control. At the “moderate” stage, the device checks after every 5 breaths whether the target volume has been reached and changes pressure by 1.0 hPa. If the pressure reaches a corridor around the target volume, the device switches to precise control. At the “fast” stage, the device checks after every breath whether the target volume has been reached and changes pressure by 1.5 hPa. If the pressure reaches a corridor around the target volume, the device switches to precise control. – Physiological alarms: triggered in the event of the alarm limit being undershot during at least three of the last five breaths; reset if the alarm limit is exceeded during at least three of the five subsequent breaths All values determined under ATPD conditions (ambient temperature and pressure, dry). The right to make design modifications is reserved. 62 EN Technical data