Chapter 10
Digital Communication Interface
The 7321 bath is capable of communicating with and being controlled by other
equipment through the digital interface. Two types of digital interface are
available - the RS-232 serial interface which is standard and the IEEE-488 GPIB
interface which is optional.
With a digital interface the bath may be connected to a computer or other
equipment. This allows the user to set the bath temperature, monitor the
temperature, and access any of the other controller functions, all using remote
communications equipment. In addition the cooling may be controlled using the
interface. To control the cooling with the interface the cooling power switch must
be ON.
10.1 Serial Communications
The RS-232 serial interface allows serial digital communications over fairly long
distances (15.24 meters). With the serial interface the user may access any of
the functions, parameters and settings discussed in Section 9, Controller
Operation with the exception of the baud rate setting. The serial interface
operates with eight data bits, one stop bit, and no parity.
10.1.1 Wiring
The serial communications cable attaches to the bath through the DB-9
connector on the back of the instrument. Figure 7 shows the pin-out of this
connector and the suggested cable wiring. To eliminate noise, the serial cable
should be shielded with low resistance between the connector (DB-9) and the
shield.
10-1