Appendices: Appendix E — Analog Output
BP Analog Output:
BP analog output is only available on Tram modules with the BP
monitoring feature.
NOTE
A buzz will be sent to the analog out if the BP is not zeroed.
„
The pressure labeled ART (BP 1-4) is sent to the analog out.
„
If two or more pressures are labeled ART, for example, BP-3 and BP4, the first (BP-3) will be sent to the analog out.
„
If all the BPs are labeled ART, then BP-1 will be sent to the analog
out.
„
If there is no pressure labeled ART, then BP-1 will be sent to the
analog out.
All BP outputs are 10 mV/mmHg when zeroed.
Digital Synchronization Output:
A digital sync output is available with a return marker on those modules
with the defib sync feature. (Refer to the module's service manual for
specifications.)
Tram-rac Housing
The analog output from the Tram-rac housing provides the following
signals when using the analog output cable. (Refer to the service manual
for the specific pin configuration.)
NOTE
The Solar ECG module does not support analog output from the
Tram-rac analog output connector. When using the Solar ECG
module, analog output is only available from the optional Defib Sync
connector on the front of the module.
ECG Analog Output:
NOTE
Refer to the Tram-rac Housing section in Chapter 2, Equipment
Overview, for location of the analog output connectors on the various
housings.
21-28
Solar 8000i Patient Monitor
2026265-001
Revision A