Download User Manual - White Bear Photonics
Transcript
3.0 General Operation Fig 3.4(a) STEP 3: The drop must be a stable sessile drop for accurate results to be obtained. The drop should stand upright on the plinth, be proud of the quartz surface and be of uniform shape. The drop must be similar to the drop shown in Fig 3.4(a). Fig 3.4(b) If the drop is similar to the drop shown in Fig 3.4(b), It must be removed and a new sample must be pipetted. NOTE: If the drops are consistently like those shown in Fig 3.4(b) it may be due to contamination of the plinth. In this case cleaning or changing of the plinth may be necessary (See section 4.0 Maintenance). Fig 3.5. STEP 6: Rotate the lid back so that it clicks into place with the light source directly over the plinth. Fig 3.5. NOTE: Do not rotate the lid more that 120˚ at one time. Always allow for the lid to click into place before moving it again. STEP 5: Use the Cary 50/60 Win UV software to take a measurement of the blank sample. Version 1.1 Drop Technology - 12 -