Download Agilent SPS 3 Autosampler
Transcript
Installation Instrument Communication Port The instrument communication port, see Figure 17, is used to connect the Agilent SPS 3 to the computer controlling your instrument. To connect your Agilent SPS 3: 1 Plug one end of the supplied RS-232 cable into the instrument communication port. 2 Connect the other end of the RS-232 cable to a serial port (COM1) on your computer or to the serial port on the Agilent 5100 ICP-OES. Ensure that the correct communications parameters are set for the Agilent SPS 3, the computer or on the 5100 ICP-OES. The required instrument communication settings are listed within Table 4. Only the Baud rate setting can be altered, via the instrument software. The instrument software must also be set to the corresponding values. Table 4. Communication parameters for Agilent SPS 3. Baud rate: 9600 Data bits: 8 Parity: None Stop bits: 1 Flow control: None I/O Port The I/O port provides an accessory interface, for example, this interface is used for communication between the Agilent SPS 3 and the Agilent SVS (Switching Valve System). Agilent SPS 3 Autosampler User’s Guide 41