Download MULTITEST - CDI Torque Products
Transcript
Calibration 5. Enter the following code into the display using the DIGIT right and left keys and the up and down VALUE keys: CODE5222 6. Press ENTER. — The LED display reads: SELECT — The LCD display reads: Calibrate CW 7. Press ENTER. — The LED display reads a flashing number in the LSD. — The LCD display read: Enter no. of transducers 8. Using the up and down VALUE keys, enter “1” for a single transducer or “4” for the 4-in-1 transducer. — If you enter "1" skip to step 11. 9. Press ENTER. The LED display reads a flashing number in the LSD and the LCD display reads: Select transducer 10. Using the up and down VALUE keys, enter the number of the 4-in-1 transducer to calibrate. Possible values are: 1 = 50 in oz 2 = 400 in lb 3 = 1000 in lb 4 = 250 ft lb 11. Press ENTER. — The LED display banners: ENTER FS followed by the last full scale calibration point for the particular transducer installed or selected with its MSD flashing. — The LCD display reads: ENTER FS transducer [units] 12. Using the up and down VALUE keys and the left and right DIGIT keys, enter the full scale torque capacity of the transducer in the engineering units selected. 13. Press ENTER. — The LED display banners: ZERO TARE Four zeros follow shifting alternately across the display. — The LCD display reads: CALIBRATION <ZEROTARE> CONT. 5-9