Download Note - Frank`s Hospital Workshop
Transcript
2-34 Chapter 2: Unpacking & Setup Operator's manual The Extended Systems Test (EST) The EST function is accessed from the Setup screen as shown here. Press the SETUP membrane button to the lower left of the UIM to open this screen. Figure 2–44: Setup Screen Press the EST touch screen button to select it. A message will appear instructing you to remove the patient and block the patient wye. After confirming that the patient has been disconnected and the circuit wye blocked press the Continue (Cont) button. The ventilator will perform the EST and display a countdown clock. During this test the ventilator will perform: • Patient circuit leak test • Patient circuit compliance measurement • Two point calibration of the oxygen sensor The patient circuit compliance measurement and leak test are performed simultaneously with the oxygen sensor calibration. The maximum time for the EST is 90 seconds. To restart the EST at any time select the Cancel button to return to the set up screen. Figure 2.46a Remove Patient Instruction Figure 2.46b In progress EST After each test is complete the ventilator will display a “Passed” or “Failed” message next to the corresponding test. Once the test is complete press the continue button to return to the set up screen. Figure 2.46c Completed EST Figure 2–45: EST Screens L2786