Download NOTE - Frank`s Hospital Workshop
Transcript
E-2 Appendix E Sensor Specifications & Circuit Resistance AVEA Ventilator Systems Hot Wire Flow Sensor Specifications Table E-2 Hot wire sensor specifications Part Number Type: Circuit Location: Performance Specifications Flow Range: Vol. Accuracy: Flow Resistance: Dead Space: Freq. Response*: Calibration: Linearity: Operating Temperature: Physical Specifications Sensor length Diameter Insp (Vent Side) Diameter Exp (Patient Side) Tube length Connector Weight Service Life Sterilization Materials 51000-40081 Multiple use heated wire Wye 0 (+/- 0.002) to 30 L/min +/-10% 15 cmH2O @ 20 L/min 0.8 mL 16 Hz 36 point curve < 2% 5 to 40oC 1.68” 15 mm OD 15 mm OD N/A Pin & Socket type < 10g (not including wire) 25 cycles Steam Autoclave Sensor - Delrin Wire – Platinum Screen – Stainless Steel 304 or 316 Pin – PhBz, gold over nickel plated Spacer - Delrin Circuit Resistance (per EN794 –1) It is important to check the inspiratory and expiratory resistance specification of patient circuits used with the AVEA to ensure they do not exceed the following limits when adding attachments or other components or subassemblies to the breathing circuit. NOTE Refer to product labeling supplied with any accessory to be added to the breathing circuit for this information. 0.6 KPA (6cmH2O) at 60 L/min for adult patients 0.6 KPA (6cmH2O) at 30 L/min for pediatric patients 0.6 KPA (6cmH2O) at 5 L/min for neonatal patients L2786