Download ECG Demo Board Final Report
Transcript
of two system bandwidths (50 Hz and 100 Hz). Another feature added to the final Phase 3 PCB was an LED indicator light to alert the user when the circuit was being powered. Adding the LED reduced the battery-life of board, however with the LED drawing only 1.5mA, the AFE final board still has a sufficient battery life of 211 hours of continuous operation. Figure 16 below shows the schematic for the phase 3 board including the additional circuitry and features. Figure 16: Phase 3 schematic (with switches) developed in PCB Artist Keeping the same sub-layout of each individual circuit element (servo loop, LDO, output filter, etc.) the layout of phase 3 was spread out to allow for labeling. The precision analog group requested this for the purpose of allowing customers an ability to easily recognize the circuitry used in the board as well as to highlight the components featured in the circuits. The thumb pads were separated to either side of the board to make it easier to grasp when the board is mounted on a display. This was done to make it easier for the user to grip the AFE board and move as little as possible (reduce DC drift and offset). Slight movements in the electrode (finger) connection were researched and found to create artifacts in the ECG signal. The inputs and outputs were also grouped together (100 mil spacing) to allow Molex connectors and custom cables 24