Download Synapse CCD Manual
Transcript
Initial Power-up and Operation • ADC Speed selection – sets the rate at which the data is read off the CCD detector. For maximum signal to noise, the ADC speed should be set to 20 kHz. For maximum frame rates, the ADC speed selection should be set to 1 MHz. • Time Interval – the elapsed time between the start of one accumulation to the start of the next accumulation. The Time Interval, Integration Time and Readout Time of the CCD detector have the following relationship: t interval ≥ t integration + t read 27