Download Solutions to Sticky Problems

Transcript
CHAPTER 3
3.1 Why You Should Read This Chapter
The purpose of this chapter is to provide the Viscometer user with information necessary to make
meaningful viscosity measurements. It will describe
the mechanical components of the Brookfield Viscometer and suggest some useful operational techniques.
Those adhering strictly to the Pragmatic school of
viscosity measurement may not wish to read any
further than this chapter. All users, however, should
read it before moving on; a good grounding in basic
Viscometer operation will facilitate advancement to
more sophisticated techniques.
3.2 How the Brookfield Viscometer Works
The Brookfield Viscometer is of the rotational variety. It measures the torque required to rotate an immersed element (the spindle) in a fluid. The spindle is
driven by a motor through a calibrated spring; deflection of the
spring is indicated by a pointer and dial (or a digital
display). By utilizing a multiple speed transmission and
interchangeable spindles, a variety of viscosity ranges
can be measured, thus enhancing versatility of the
instrument.
For a given viscosity, the viscous drag, or resistance
to flow (indicated by the degree to which the spring
winds up), is proportional to the spindle’s speed of rotation and is related to the spindle’s size and shape
(geometry). The drag will increase as the spindle size
and/or rotational speed increase. It follows that for a
given spindle geometry and speed, an increase in viscosity will be indicated by an increase in deflection of
the spring. For any Viscometer model, the minimum
range is obtained by using the largest spindle at the
highest speed; the maximum range by using the smallest spindle at the slowest speed. Measurements made
using the same spindle at different speeds are used to
detect and evaluate rheological properties of the test
fluid. These properties and techniques are discussed
in Chapters 4 and 5.
The Viscometer is composed of several mechanical
subassemblies. See Figures 3-1 for a schematic view
of the major components of a basic dial-reading Viscometer.
The drive motor and multiple-speed transmission are
located at the top of the instrument inside the housing
to which the nameplate is attached. The Viscometer
main case contains a calibrated beryllium-copper
spring, one end of which is attached to the pivot shaft,
the other end is connected directly to the dial. This
dial is driven by the transmission and in turn drives the
pivot shaft through the calibrated spring. In dial-reading models, the pointer is connected to the pivot shaft
and indicates its angular position in relation to the dial.
In Digital models, the relative angular position of the
MORE SOLUTIONS TO STICKY PROBLEMS
pivot shaft is detected by an RVDT (rotary variable displacement transducer) and is read out on a digital display.
SYNCHRONOUS
MOTOR
HOUSING
GEAR TRAIN
CLUTCH
DIAL
POINTER
CALIBRATED
SPIRAL SPRING
PIVOT SHAFT
PIVOT CUP
JEWELLED
BEARING
GUARDLEG
SPINDLE
SAMPLE
CONTAINER
Figure 3-1
Below the main case is the pivot cup through which
the lower end of the pivot shaft protrudes. A jewel
bearing inside the pivot cup rotates with the dial or
transducer; the pivot shaft is supported on this bearing by the pivot point. The lower end of the pivot shaft
comprises the spindle coupling to which the
Viscometer’s spindles are attached.
3.3 Viscosity Measurement Techniques
As with any precision instrument, proper operating
techniques will improve effectiveness of the Brookfield
Viscometer. A step-by-step procedure for Viscometer
operation can be found in the Instruction Manual supplied with each unit, and is not repeated here. Instead,
we present recommendations and advice gleaned from
over 65 years of customer experience. They form a
sound foundation for a viscosity testing procedure and
a starting point from which more advanced techniques
can be explored.
3.3.1 Record Keeping
We recommend that the following information
always be recorded when making a viscosity measurement; viscometer model, spindle (or accessory), rotational speed, container size or dimensions,
sample temperature, sample preparation procedure
(if any), and whether or not the spindle guardleg
was used. Test Report Forms supplied in the instruction manual with each Viscometer are convenient for this purpose.
Page 8
Brookfield Engineering Labs., Inc.