User Manual Appendix RESULTS: Total number of patients: HeartSine Samaritan AED = 8 Biphasic “X” = 23 VF Summary HeartSine Samaritan AED Parameter Biphasic “X” % % Total number of episodes 28.0 100.0 62.0 100.0 Total number of successful shocks 28.0 66.7 61.0 39.4 Total number of unsuccessful shocks 14.0 33.3 94.0 60.6 Success within 1 shock 18.0 64.3 36.0 58.1 Success within 2 shocks 24.0 85.7 43.0 69.4 Success within 3 shocks 28.0 100.0 48.0 77.4 Average number of shocks required 1.5 Average energy required (J) 125.0 Average energy reduction (J) 25.0 2.5 150.0 16.6 SUMMARY The initial results from the Human Clinical trial suggests that the HeartSine SCOPE biphasic waveform is at least as effective as the other marketed, constant energy biphasic waveform system. HeartSine Technologies, Inc Page 67