Download User Manual - AED Superstore

Transcript
User Manual
Appendix
RESULTS:
Total number of patients:
HeartSine Samaritan AED = 8
Biphasic “X” = 23
VF Summary
HeartSine
Samaritan AED
Parameter
Biphasic
“X”
%
%
Total number of episodes
28.0
100.0
62.0
100.0
Total number of successful shocks
28.0
66.7
61.0
39.4
Total number of unsuccessful shocks
14.0
33.3
94.0
60.6
Success within 1 shock
18.0
64.3
36.0
58.1
Success within 2 shocks
24.0
85.7
43.0
69.4
Success within 3 shocks
28.0
100.0
48.0
77.4
Average number of
shocks required
1.5
Average energy required (J)
125.0
Average energy reduction (J)
25.0
2.5
150.0
16.6
SUMMARY
The initial results from the Human Clinical trial suggests that the
HeartSine SCOPE biphasic waveform is at least as effective as the other
marketed, constant energy biphasic waveform system.