Download PQA300 Picture Quality Analysis System User Manual
Transcript
Appendix C: Functional Check 13. Give the test capture an alias name that helps you remember the correct capture and reference to associate it with. For example, NTSC Notch Mobile Test or PAL Notch Mobile Test. 14. Click OK to accept the changes and exit the menu. 15. Click the Capture soft key to capture the Mobile+ sequence as a test. 16. Click the Measure soft key to start the measurements. The picture quality is analyzed with respect to the captured reference. The effects of the encoder and decoder modules are eliminated (de--embedded) from the analysis. 17. Check the Summary results against the typical values for the unimpaired video test sequence given in Table C--5. The values you get for the PQR and PSNR numbers can be different, but should be about the same. Table C- 5: Summary picture quality values for unimpaired sequence compared to a captured composite reference using Notch processing Composite Impaired Sequence Test Test NTSC and NTSC no setup PAL PQR Luma 0.90 1.40 PQR Luma & Chroma 0.90 1.40 PSNR Luma 48.00 45.00 When the unimpaired composite video test is correct, use the following procedure to test impaired video sequences. Measuring an impaired test sequence against the stored reference sequence checks the operation of encoder and decoder modules with different levels of impairment in the video material. 1. In the Generate Configuration menu, select one of the impaired video test sequences (for example, Mobile3mbs+). 2. In the Capture Configuration menu, select the correct signal type for the test (NTSC, NTSC No Set Up, or PAL). 3. Select Mobile+ as the reference sequence. 4. Set Capture As to Test. Give the test capture a meaningful alias name so you can identify it in a list of captured files. That name is also displayed on the results displays to identify the results. 5. Click on the Decoder soft key. 6. Select Adaptive Line processing. Adaptive Line should be used in most cases when testing against the 601 digital reference. 7. Click OK to accept the changes and go to the Capture Configuration menu. C- 18 PQA300 Picture Quality Analysis System User Manual