Download RT-ZD10/20/30 User Manual

Transcript
R&S®RT‑ZD10/20/30
R&S RT‑ZA15 External Attenuator
Maximum Input Voltage
3.5
Maximum Input Voltage
The maximum input voltage of the R&S RT‑ZA15 external attenuator is specified
as the maximum voltage between each signal socket and ground.
Shock hazard caused by high voltages
To avoid electric shock and personal injury, and to prevent damage to the
probe or any other products connected to it, observe the following instructions:
● The maximum input voltage for DC signals is ±70 V between each signal socket and the ground socket.
● The maximum effective input voltage for AC signals is 33 V (eff)
between each signal socket and the ground socket. The maximum transient peak voltage is ±46 V.
The maximum input voltage is derated for higher frequencies. Refer to the
R&S RT‑Z data sheet for further information.
User Manual 1410.4550.02 ─ 03
34