Algorithm Flow Diagram for Visibility Sensor Sample the forward scattered light thirty seven (37) times each second in Phase I and Phase II. Sample the illuminating LED power thirty seven (37) times each second as Pwr. || Subtract Phase II from Phase I to generate a one (1) second RcvrAvg || Add the one( 1) second average to the Painting Average || Adjust the average for power: AvgRcvr=RcvrAvgXTxPwrCal Pwr Where TxPwrCal is a calibration value || Convert to Extinction Coefficient (ExCo) in inverse kilometers as AvgExCo=f(AvgRcvr, Span, Zero) AvgExCo available for polling by the DCP >>>>> >>>>>>>>>>>>>>>>>>>>>>>> >>>>> Where DCP polling command is “FL” Where Span and Zero are calibration values From ALS, one (1) minute average ALS data available for polling by the DCP Where DCP Polling command is “FE”