Download User`s Guide DSA800 Series Spectrum Analyzer
Transcript
Chapter 2 Front Panel Operation RIGOL TG Lvl Offset Assign a certain offset to the output power of the TG when gains or losses occur between the TG output and external device in order to display the actual power value. Key Points: This parameter only changes the readout of the TG output power, rather than the actual value of it. The offset could be either a positive (gain in the external output) or a negative (loss in the external output). You can use the numeric keys, knob or direction keys to modify this parameter. For more details, please refer to “Parameter Setting”. Table 2-22 TG Level Offset Parameter Explanation Default 0 dB Range -200 dB to 200 dB Unit dB Knob Step 1 dB Direction Key Step 10 dB Power Sweep Enable or disable the power sweep function. When enabled, the output power of the TG varies with the sweep rate of the analyzer within the specified sweep frequency range (from start to stop frequency) and increases gradually from the specified TG power (see “TG ”) within the specified power range (see “Power Range”). When disabled, the TG output is fixed at the specified power (see “TG ”). Note: this function in not available in zero span mode. User’s Guide for DSA800 Series 2-41