Download file attachment
Transcript
Installation Instructions Detailed instructions are included in Chapter 4 of the Model 2100 User’s Manual (Document Number: 2100-900-01B). The manual is available on the Product Information CD that ships with the Model 2100 Multimeter and is also available for download from the Keithley website at http://www.keithley.com/support/data?asset=50760 . Alternatively, Keithley factory upgrades can be arranged by calling your local Keithley support office. v1.08 Release Overview v1.08 is a maintenance release. Critical Fixes Symptom: The 2100 will hang when the sample count is set over 32 and a “read?” query is sent. Resolution: Issue has been resolved. When the sample count is set over 32, the 2100 will now successfully execute a “read?” query. Non-Critical Fixes Symptom: When the language of the 2100 is set to A34401, the settings for NPLC, Remote Digits Displayed, and Resolution remained set to Default values. Resolution: Issue has been resolved. When the language of the 2100 is set to A34401, the instrument settings for NPLC, Remote Digits Displayed, and Resolution are set at values equivalent to that of the Agilent 34401A as seen in the tables below. Integration Time(PLC) DEFAULT A34401 0.02 0.02 0.1 0.1 Page 2 of 8 Auto Zero Resolution Remote Digits Displayed DEFAULT A34401 DEFAULT A34401 Fast 4 _ digits Off Off 4_ 4_ 1 Slow 4 _ digits Off On 4_ 5_ 0.1 Fast 5 _ digits Off Off 5_ 5_