Download User Manual
Transcript
3. 4. 5. Raise the high voltage in uniform steps of 50 volts. At each step note the radiation intensity on the CPM/S range. Do not go above the maximum voltage for the detector detector. Plot the High voltage vs. counts on a graph. Adjust the high voltage in the instrument to the value that corresponds to midway on the plateau. If there is no upswing, then adjust it above the knee. SET HIGH VOLTAGE TO MIDPOINT KNEE 1000 COUNTS f) The other way is to use the Custom Settings Menu which allows programming into the locations in EEPROM. This is more complicated and involves looking up the locations of each item in the list. We recommend using the Detector Setup Menu. Remember that the Detector that is selected using the Function Menu will be the detector whose parameters you are saving or changing. See the individual sections on how to use these Menus. If you have problems with the programming and the instrument is functioning in an unpredictable manor, you can revert back to the factory settings under the Setup Menu, however the values for the previous calibration will be lost. g) UPSWING Programming the Setup Value The values on the worksheet may be programmed into the instrument in either of 2 ways. The first and easiest way is to use the Detector Setup Menu. This is an interactive mode that shows the count rate as well as the settings for each of the above items. Log Bargraph If you have a Log Bargraph the Range Low Limit sets the low end of the bargraph. Set the Range High Limit as you would if using a normal bargraph The with the log bargraph, the digital display will change ranges at each decade. PLATEAU 0 0 1050 HIGH VOLTAGE GRAPH OF HIGH VOLTAGE VS COUNTS (2) Discriminator The Discriminator is set by lowering the value to the point where the instrument will count from noise, then raising it past the point where it stops counting. Check for double counting at low count rates. Lower the gain or raise the discriminator to correct this problem. (3) Saturation The saturation detector if installed, is used to prevent a saturated detector from reading 0 on the display. Set it to 0 to turn it off. To adjust the setting in Detector Setup Mode: 1. Reduce the Saturation to zero. 2. Expose the detector to sufficient radiation so the reading is at the top of the highest scale. 3. Raise the saturation until the reading shows off scale. 4. Lower the saturation by 10%. e) Other variables and settings The bargraph type, display type, inverted or normal display, modes for each detector, and range up and down times can be reviewed under the Custom Settings section. MODEL 5085 MERIDIAN Health Physics Instruments 19