Download INSTALLATION AND OPERATION MANUAL
Transcript
Main menu Page 21/46 6. Main menu Following menu structure is available: Meter setup Measure Outputs Communication Misc Info PIN Meter Setup Calibration This figure is used for setting pipe’s diameter (size). Several sizes from DN 15 to DN 600 can be set. Diameter [Diameter] Note: Pipe diameter is set at the factory. Changes of size have an impact on meter’s accuracy. U This parameter is set at the factory. This factor compensates for accuracy error as a result of the installed detector. If accuracy adjustment of the meter is required, please refer to the scale factor. Detector Factor [Det Fact] A In the event the amplifier is replaced, this parameter must be reprogrammed with the original detector factor. This parameter is set at the factory. This factor compensates for accuracy error as a result of the installed detector. If accuracy adjustment of the meter is required, please refer to the scale factor. Detector Offset [Det Zero] A This parameter is set at the factory. This factor compensates for accuracy error as a result of the installed amplifier. Amplifier Factor [Amp Fact] Read only Coil Current [Coil Cur] Read only This parameter is set at the factory. This factor compensates for accuracy error as a result of the installed amplifier. MID_M5000_BA_02_1304