Microbiology Reader Bioscreen C User's Manual ____________________________________________________________________________________________________________ Figure 3 Placing the filter wheel. Connect the power cord to the back of the Bioscreen instrument. Plug it in. Caution: Before turning the Bioscreen on ALWAYS remember to check that the incubator cover is in its right position. Improper placing of the cover can cause serious damage to the instrument which is NOT covered by warranty! Always close the lid before starting an experiment, as placing fingers etc. inside a running instrument can cause serous physical injury. Please see chapter 4.4 (Start up) for more information. Also remember to check that the green cover of the measurement compartment is properly in place. Turn on the instrument (the switch is on right side towards the back). 2.2.2 Running the instrument test: • When the instrument is ready to start, 0.0 is shown on the LED display. • Press MODE on keypad, display will read tESt. • Press 21, then press ENTER on keypad and the pump will start etc. • Allow the instrument to run for at least 5 minutes. • If there is a problem, the test sequence will automatically terminate. • If the instrument is OK, it will keep displaying readings and continue to run. • To stop the instrument, press the RESET key, then switch the instrument off. ____________________________________________________________________________________________________________ 4