Transcript
2.2 Installation Location for the Instrument Install in a location that meets the following requirements: CAUTION (1) Instrument panel or rack This product should be mounted on an instrument panel or in a rack. (2) Well-ventilated location To keep instrument cool, mount in a well-ventilated location (3) Location with minimal vibration Install in a location with minimal vibration. Instrument accuracy may be affected if the installation location is subject to vibration. (4) Horizontal location As far as possible, the installation location should be horizontal. Avoid locations with the following conditions: CAUTION (1) In direct sunlight or near hot objects Choose an installation location with stable temperature (around ambient 23°C). Placing this instrument in direct sunlight or near hot objects may adversely affect accuracy. (2) Location with oily smoke, warm or humid air, dust, corrosive gas Avoid such locations. (3) Near a source of electromagnetic fields Electromagnetic fields may affect the accuracy of the instrument. 2-4 IM 11M10A01-01E