Download Silver Writing4
Transcript
‘‘The Silver Sheet’’ Elsevier Business Intelligence Manufacturer Product Class Reason Smith & Nephew Peri-Loc II Inner packaging of the sterile product was not sealed. (Z-2314-2010) Smith & Nephew Cerclage Wire II Plastic trays that encase these products have burrs on the hinged packaging, which can puncture the single-barrier sterile pouch, compromising sterility. (Z-2197/2199-2010) Smith & Nephew Aimer Endofemoral II Tip may separate due to failure of the soldered joint. (Z‑0040/0048-2011) Smith & Nephew R3 22 MM I.D. II The femoral head may pop out of the liner intraoperatively. (Z‑0079/0086-2011) Smith & Nephew Trigen Sureshot Distal Targeting System II Software has been found to provide erroneous targeting. (Z‑0411-2011) Smith & Nephew Fast Fix 360 Needle Delivery System II System may not allow the T-2 anchor to advance for implantation. (Z-0419/0421-2011) Smiths Medical Medfusion Syringe Infusion Pumps II Pumps with obsolete software may continue to run past the set volume limit. This can result in over-delivery if the syringe is overfilled. (Z-2121-2010) Smiths Medical AC Power Cords II Potential for the power cord’s prongs, made by Electri-Cord Manufacturing, to crack and fail at/or inside the plug. (Z-1082-2010) Smiths Medical CADD-Legacy 1 Ambulatory Infusion Pump II Pump was shipped without documentation that required accuracy testing had been performed during the manufacturing process. If the pump did not go through accuracy testing and does not meet accuracy specifications, there is a potential for over- or under-delivery. (Z-1235-2010) Smiths Medical Medfusion Syringe Infusion Pumps II There has been an increased trend in reports of “Motor Not Running” (MNR) and “Motor Rate Error” (MRE) alarm-message events. This is likely to cause an interruption or delay in therapy for the patient. (Z-0650/0651-2010) Smiths Medical Power Cords II According to FDA, two medical device manufactures have reported incidents of sparking, charring and fires from power cords made by Electri-Cord Manufacturing. (Z-0918/0920-2010) Solafeet Foot Portable Tanners II Product failed to have a label that contained a recommended exposure schedule, and the operator’s manual provided inadequate instructions for use. (Z-0608-2010) SonaMed Clarity System II II There is insufficient evidence to support any claims for safety and efficacy. (Z-2165-2010) Sonosite Tee Transducer II The transducer, when connected to a Sonosite M-Turbo ultrasound system, may exhibit erratic operation of the “Scanplane Orientation Indicator” on the system display. The indicator may move erratically and not accurately represent the orientation of the Tee transducer array. (Z-2191-2010) Sonosite PowerPark System II System has the potential for loss of power at the corresponding wall outlet and/or circuit due to poor alignment. (Z-2116-2010) Sonosite ARM Software III Software on the NanoMaxx ultrasound system results in errors when the “Print” command is selected. As a result, the displayed MI or TI values may be incorrect. (Z-0478-2011) Sorin Group Paradym CRT-D Devices II Devices have a software anomaly that causes the device to lose the ability to sense/pace and deliver further therapy. (Z-2579-2010) 50 | FEBRUARY 2011 © 2011 F-D-C Reports, Inc., an Elsevier company. All rights reserved.