Transcript
Recommendations • It is recommended that wherever possible ISO standard symbols with text be used rather than using the Main manufacturers’ ownDisplay symbols. Where the required symbol is not available as a standard, the manufacturer should seek to influence the ISO standard to include a new symbol. • Symbols need to be tested with users in context. There is an ISO standard for developing and testing symbols with users.19 • It is recommended that manufacturers Volume add text labels to symbols History until the meaning of the symbols are universally recognised by users. • Symbol recognition and understanding of function should be incorporated into training and competency assessment materials. The on/off button should be the only button with 2 functions. In line with MHRA requirements, turning off the pump should be a 2 step process requiring confirmation from the user. start Use green as an additional indicator for the start button stop Use red as an additional indicator for the stop button Bolus Use 3 arrows to indicate the bolus function and avoid confusion with rewind / fast forward icons. Keep icons simple and minimalist. Remove any unnecessary graphics where possible. P Where abstract icons are used, use labels to help identify the button. A guide to the design of electronic infusion devices |33