Transcript
Performance Characteristics The minimum detectable dose of C8 as calculated by 2SD from the mean of a zero standard was established to be 0.4 ng/ml. Intra-assay precision was determined by testing replicates of three plasma samples in one assay. Inter-assay precision was determined by testing three plasma samples in twenty assays. Sample n CV (%) Average CV (%) Intra-Assay Precision 1 2 3 20 20 20 4.6% 4.3 % 3.9% Inter-Assay Precision 1 2 3 20 20 20 7.5% 8.1% 7.2% 4.3% 7.6% Recovery Standard Added Value Recovery % Average Recovery % 1.5 – 15 ng/ml 88 – 111% 103% Linearity Plasma and serum samples were serially-diluted to test for linearity. Average Percentage of Expected Value (%) Sample Dilution Plasma Serum 1:5000 96% 94% 1:10000 100% 99% 1:20000 106% 107% Cross-Reactivity Species Canine Monkey Mouse Rat Swine Rabbit Bovine Proteins Cross Reactivity (%) None <20% None 1% None None None Cross Reactivity (%) 7