Conductivity Calibration
To start the conductivity calibration, It is necessary to connect the conductivity cell and the temperature sensor
to the instrument, and then press
for 3 seconds.
CAL
1
Rinse the conductivity cell with distilled
water and press
Rinse electrode and press
CAL
Immerse the cell in a standard
conductivity solution (for example 1413
2
µs/cm), and wait for the stable reading.
1413
When
appears , press
.
Press
MEAS
1413
3
When calibration is finished, the
µS/cm
21.6 °C ATC
instrument return automatically in
measurement mode.
To make a multipoint calibration repeat
all the steps for every interested point.
1413
This instrument can be calibrated automatically on these points:
 84,0 µS – 1413 µS – 12.88 mS – 111,8 mS