Easy is the name of the game Conductivity Calibration To start the conductivity calibration, It is necessary to connect the conductivity cell and the temperature sensor to the instrument, and then press for 3 seconds. CAL 1 Rinse the conductivity cell with distilled water and press Rinse electrode and press CAL Immerse the cell in a standard conductivity solution (for example 1413 2 µs/cm), and wait for the stable reading. 1413 When appears , press . Press MEAS 1413 3 When calibration is finished, the µS/cm 21.6 °C ATC instrument return automatically in measurement mode. To make a multipoint calibration repeat all the steps for every interested point. 1413 This instrument can be calibrated automatically on these points: 84,0 µS – 1413 µS – 12.88 mS – 111,8 mS