Download Professional Plus User Manual

Transcript
The pH calibration allows up to a six point
calibration. Once the reading is stable,
highlight Accept Calibration and press enter
to accept the displayed Calibration value or
highlight Calibration value and press enter
to input a new Calibration value. Once
this value is accepted, the message line will
display “Ready for 2nd Point” and you can
move to the next point if needed. Press Cal
to complete the calibration or press Esc
to cancel the calibration. The instrument
has auto buffer recognition available so the
Calibration value is likely to be accurate if
the correct buffer set has been selected in the
setup menu.
Place the sensor in the second solution, and
allow the reading to stabilize. The message
line shows the instrument is “Ready for 2nd
point” calibration value. Once the readings
are stable and the instrument has determined
the buffer value, verify Accept Calibration
is highlighted and press enter to confirm the
second calibration point.
38