Download User Manual - Parts Center

Transcript
Multi-function Flue Gas Analyser
ANTON
Tel: +44 (0) 870 428 0075
Fax: +44 (0) 870 428 0076
[email protected]
www.anton-group.com
Anton Group
Tel: +44 (0) 870 428 0075
V2.2 i1.05
User Manual
M071021
January 2013
Issue 2
Summary of Screen Icons
þ
é
ê
ç
è
P=0
u

¢tt
ü
û
Select, pick or action
Up item or entry
Down item or entry
Left field
Right field
Zero pressure
•
•
•
•
•
Start
(Pressure & CO Room Safety tests)
Stop
(Pressure & CO Room Safety tests)
Restart
(Restart timer)
Yes, pass or done
No or fail or cancel
Purge icon
Hold / pause readings
T1
T2
Select temp 1 (flow)
T1
T2
Select temp 2 (return)
<
Safety information:
(Differential Temperature test)
(Differential Temperature test)
Save log
Print
Additional information:
Sprint is designed to support the working practices defined
in British Standard BS7967 and the Design Standards
EN50379, BS7927 for flue gas analysers. It is highly
recommended that users are fully conversant with BS7967
when using a flue gas analyser for servicing or installing a
boiler system.
Sprint offers a timed let-by/tightness test in accordance
with the UK's Institute of Gas Engineers' procedure IGE/11/
UP/1B.
Performing CO measurements
When performing any CO measurements ensure the unit
is zeroed in clean air in accordance with British Standard
BS7967. A suitable location for sampling clean air will be
outside of the building where the boiler system is installed.
1. Repair of this equipment and gas sensor replacement
shall be carried out by the manufacturer or certified
service centre in accordance with the applicable code
of practice.
Send XML report to Bluetooth®
(eVo3 only)
Serial communications with PC in progress – please wait
>
>
––
+
–
+
–
Darker contrast
2. If the equipment is likely to come into contact with
aggressive substances, then it is the responsibility of the
user to take suitable precautions that prevent it from
being adversely affected.
Lighter contrast
3. The equipment is designed for use in ambient tempera-
Delete
Restore default contrast
Edit value
Increment value
Decrement value
Decrement through character list
(String edit)
Increment through character list
(String edit)
Logging busy
tures in the range -10°C to +50°C and should not be
used outside this range
4. Use only charger(s) supplied by Telegan/Anton.
5. Use only the appropriate Telegan/Anton supplied cables
for connection to the sockets on the instrument.
© Copyright Telegan Gas Monitoring 2013.
All rights are reserved. No part of the document may be photocopied, reproduced, or translated to another language without the
prior written consent of Telegan Gas Monitoring
Publication number: M071021
Second edition: January 2013
CO Alarm
η
Read and understand all instructions in the operation
section of this manual before use.
Do not substitute components as this may impair safety
and invalidate warranty.
Observe all warnings and instructions marked on the
unit and within this manual.
If this product is not working properly, read the troubleshooting guide or call Anton.
Ensure qualified service personnel change sensors and
provide maintenance and calibration.
Select Net/Gross/NetHE Efficiency
Bluetooth® is a trademark owned by Bluetooth SIG, Inc.
and licensed to Crowcon Detection Instruments Ltd
Contents
Sprint eVo Multi-function Flue Gas Analyser..........................2
Unpacking...............................................................................................2
Overview.................................................................................................3
Probe connections...................................................................................4
Quickstart Guide....................................................................5
I. Operation.........................................................................14
1.1 Menu and operator button overview...............................................14
II. Setting Up ......................................................................16
2.1 Configuration Options ....................................................................16
III. Battery Charging.............................................................19
Charging the batteries...........................................................................19
IV. Maintenance and calibration...........................................20
4.1 Unit.................................................................................................20
4.2 Water trap.......................................................................................20
V. Specification....................................................................21
VI. Accessories and spare parts............................................23
Accessory list.........................................................................................23
VII. Logging.........................................................................24
VIII. Printing and Data Transfer.............................................25
IX. Troubleshooting guide....................................................26
Appendix I: Carbon Monoxide Room Safety Tests................28
Unpacking
Sprint eVo1/eVo2/eVo3
Sprint eVo1, eVo2, eVo3
Multi-function Flue Gas Analyser
Thank you for purchasing the Sprint eVo1, eVo2, eVo3 Multi-function Flue Gas
Analyser. Sprint has redefined flue gas analysis and will give you years of unparalleled service and reliability.
There are three versions covered in this manual as follows:
Base model flue gas analyser
Sprint eVo1
As eVo1 with pressure and GEP capability
Sprint eVo2
As eVo2 with Bluetooth
Sprint eVo3
Please read the instructions carefully before use. Keep the manual for future
reference.
Unpacking
Important: ensure unit is fully charged.
Remove the Sprint unit from the packaging. The Sprint accessories will be
located in the carry case. Check the contents are complete, you should have:
• Carry case;
• Sprint unit;
• Mains battery charger power supply;
• Flue probe, including water trap;
• A5 user guide;
• User manual on CD;
• Certificate of calibration;
• 2 x 1 m tubes for pressure (Natural gas).
Options to include:
• Gas leak probe;
• Additional thermocouple probes;
• In-car charger;
• Infra red printer.
Warning: Do not attempt to use any other charger power supply, with this
unit except the one(s) supplied. Failure to comply could invalidate the warranty and may result in permanent damage to the unit.
2
Sprint eVo1/eVo2/eVo3Overview
Overview
ANTON
Tel: +44 (0) 870 428 0075
Fax: +44 (0) 870 428 0076
[email protected]
www.anton-group.com
Soft keys
Use these unmarked buttons
to make selections from the
menu, start and stop tests,
print and store results.
Anton Group
Tel: +44 (0) 870 428 0075
V2.2 i1.05
ON/OFF/ENTER key
Press and hold button for two
seconds to switch on and switch
off Sprint unit. Use this key to
finish edits, save changes and
view next page of multiple
paged screens.
ESCAPE key
Use this button to quit
tests and step back
through menus.
(-)
Battery icon
Three bars show
complete charge
Purge icon
Rotating icon shows
pump is purging
See section Operation for more
details
Pressure inlets (eVo2
and eVo3 only)
Display
Showing start-up
splash screen
Thermocouple
sockets
K-type connectors
(+)
Power jack
Flue sample inlet
IR window
Printer output
Note: eVo3 has Bluetooth® communications
USB connector
Connect gas leak
probe here.
Also used for
communications to PC
(eVo2 and eVo3 only)
Exhaust exit
Warning: do not
inhale exit gas exhaust
or block exit
3
Sprint eVo1/eVo2/eVo3
(-) PRESSURE
(DRAUGHT)
FLUE SAMPLE
Exhaust exit
(+) PRESSURE
TEMPERATURE
Tip: see label
underneath unit
for details of probe
connections. Sprint
eVo1 does not have
prssure inlets
INLETS
Probe connections
FLOW
(EFFICIENCY)
RETURN
Probe connections
Flue gas analysis
Connect the flue probe to the flue sample inlet and the thermocouple to the
k-connector marked FLOW (Efficiency).
To measure flue draught pressure connect pressure tube to (-) Pressure inlet
marked draught. (eVo2 and eVo3 only)
Differential pressure, working pressure, operating pressure and let-by
and tightness test (eVo2 and eVo3 only)
Connect tubing to pressure inlet(s).
Differential temperature test
Connect one (or two) thermocouple probes to the k-type connectors. When
using one probe, Sprint will display a soft key option to switch between T1 and
T2 snapshot measuring points.
Room CO safety test
Connect CO room safety test probe (optional), where suitable, to the flue sample inlet.
Gas escape test (eVo2 and eVo3 only)
Connect gas leak probe to the USB connector.
Note: The pump may operate at different speeds or switch off depending on the test
being performed. This will vary the pitch of the sound from the pump and does not
indicate the pump is performing incorrectly.
Accessories
A magnet on the reverse of the instrument can be used to place the Sprint unit
in location on the boiler system for easy hands-free operation. WARNING:Take
care not to place items which may be sensitive to strong magnetic fields near this
magnet, eg credit cards or magnetic storage devices like computer hard drives.
4
Sprint eVo1/eVo2/eVo3
Quickstart guide
Quickstart Guide
Switch On
Important: Please connect flue probe first before switching on your
Sprint unit, ensure the water trap is empty and is fitted in the right
direction. Do not insert probe in flue until after auto zero. Always start
in clean air. The gas exhaust outlet is not blocked.
Press and hold the ON/OFF/ENTER key for two seconds. Sprint will emit a few
rising beeps, performs a screen test and takes you directly to the Test Menu.
The instrument can now be used for Pressure and
Main Menu
Temperature functions. When entering Flue Gas Analysis
14:44:27
08-Feb-2006
17-Jan-2007
Cal due
or CO Room Safety functions are selected the pump
switches on to purge any residual gas from the unit and
Test Menu
carry out a zero. On stabilisation of the instrument the
Fuel Options
Units of Measurement
function can now be used.
Anayser Settings
Tip: see soft key
icon list on the
inside front cover
of this manual
Stored Logs
Auto Zero
When entering Flue Gas Analysis or CO Room Safety options from the Test
Menu, Sprint will perform an Auto Zero. Ensure you are in clean air before
proceeding.
Tip: Auto zero Sprint outside of the building or well away from the heating appliance to avoid any
potential gases in the vicinity affecting the auto zero process.
Press the þ key to confirm you are in clean air. The Sprint eVo will then perform an auto zero. Provided the auto zero is successful the instrument will
display the menu option chosen.
Note: Sprint will switch off the pump when not performing tests or purging.
5
Quickstart guide
Sprint eVo1/eVo2/eVo3
Using the menu
From the Test Menu press the ESC key to display the Main Menu. Use the soft
keys to scroll, select and edit menu items (see icon list - inside front cover). Press
accept þ to make a change or the ESC key to cancel. Press the ESC key to return
to the Main Menu.
Changing fuel, units and efficiency options
From the Main Menu select Fuel Options or Units of Measurement.
Fuel Options: select one of the following options: natural gas, LPG, heavy oil,
light oil, coal, wood, coke, Biomass, Bagasse and wood pellet dry.
Units of Measurement: select one of the following options:
Pressure units (eVo2 and eVo3 only): mbar, Pa, hPa, kPa, PSI, inWG,
mmWG, inHG or mmHG
Temperature scale: Centigrade or Fahrenheit
Efficiency: Net, Gross or NetHE*
*NetHE may be used for High Efficiency/Condensing boilers
NetHE not available on all fuel types
See section II. Setting up for more details.
Changing display and key pad settings
From the Main Menu select Analyser Settings and one of the following options:
Auto off timeout: use the soft keys to adjust or disable the timeout period.
Back light: use the soft keys to select one of the following options: off, dim or
bright.
Key click: use the soft keys to enable or disable audible key click.
Supervisor Settings: See below.
Changing date & time, report header and password
From the Main Menu select Analyser Settings>Supervisor settings: (if the
password has been set, enter the password now) select one of the following
options:
Set date & time: use the soft keys – and + to change units of date and time
and è to select hours, minutes, day, month and year. Press ESC to save.
Edit report header: use the soft keys é and ê to select header text one or two
and +
– to edit text. Use the and keys to scroll through character lists
6
Sprint eVo1/eVo2/eVo3
Quickstart guide
and character values. Press è to move on to the next letter in the header
text. Press ON/OFF/ENTER to delete all characters to the right. Press è to
move the cursor to the end of the text and þ to accept change and return
to Edit report header screen. Press ESC to exit edit without saving changes.
Print Cal Due: Enable or disable printing of calibration due on reports.
Flue CO Alarm: During flue gas analysis a carbon monoxide alarm can be set
to activate at 300ppm of C sword: press the soft key +
– to edit password.
Use the
and
keys to scroll through character lists and character values. Press è to move on to the next letter in the password. Press è to
move the cursor to the end of the text and þ to accept change and return
to Password screen. Press ENTER to store the change.
Retrieving stored logs
From the Main Menu select Stored logs and one of the following options:
Select a log: use the soft keys to scroll and select log.
Find a log by number: use the soft keys to scroll up and down the numbered
logs database.
Delete all logs: this option will delete all stored logs. Press accept þ 'Are you
sure?' to clear logs or press ESC to cancel.
Switch off
Turn off unit in clean air and ensure any gas is purged from sensors. Press and
hold the ON/OFF/ENTER button for approximately two seconds. The power off
screen will be displayed and the pump will run to purge the sensors. The pump
will run for up to 30 or 40 seconds to purge unit if gas is present. Sprint will
normally switch off in 10 seconds. Press the ESC key to cancel the switch off
sequence.
Charging battery
Plug the supplied charger into the charger socket. The batteries will recharge in
four hours from flat. You may leave the unit on charge for longer periods, e.g.
overnight, without damaging the unit. A fully charged unit will give up to nine
hours of operation. A shorter charge time can be applied, such as 1/2 hour,
though it will give limited length of operation from that charge.
7
Sprint eVo1/eVo2/eVo3
Performing tests
Performing tests
From the Main Menu select Test Menu. Refer to section III Operation for more
details.
Warning: During testing, ensure the combined filter and water trap is not blocked or
full. Failure to do so may result in an error message.
1. Flue gas analysis
Before performing the flue gas analysis test, check the water trap is clean and
is upright (arrow should point in direction of gas flow). To begin test, select Flue
gas analysis from the Test Menu.
1
Sprint will switch carry out an auto
Flue Gas
14:44:27
08-Feb-2006
zero then if successful will begin
Natural gas
Tip: The screen
making measurements. Check the
15
CO
number is
ppm
fuel type displayed on the screen is
displayed here.
CO2
4.8
correct. Use the soft keys to log or
%
print the results.
Ratio
0.003
Tip: If # symbol
Pressure
Press the ON/OFF/ENTER button to
is displayed mBar
0.00
flue probe is
display the three screens available:
P=0
Screen 1: CO, CO2, CO/CO2
ratio. Pressure (eVo2 and eVo3
only)
Screen 2: O2, excess (XS) air,
ESC
temperature and efficiency.
Final screen: combines elements from previous screens.
not in flue or
not connected.
Tip: Press the
ON/OFF/ENTER
button to cycle
through the
screens.
To perform a pressure zero (eVo2 and eVo3 only), use the soft key marked
on the relevant screen
To scroll through the efficiency options, use the soft key marked η.
Screen 1
8
Screen 2
Screen 3
P=0
Sprint eVo1/eVo2/eVo3
Performing tests
Note: for personal safety a carbon monoxide (CO) alarm will activate at 300ppm. This
will deactivate when CO levels drop below 150ppm. This is to protect the user from
potential hazardous exhaust gas levels.
To end test press the ESC key.
2. Let-by and Tightness test (eVo 2 and eVo3 only)
Let-by test
Before carrying out the test, the pressure must be zeroed with the tube connected to the instrument but not the pressure source. Note: in these tests 'Diff
press' is the difference between start and finish pressures.
1. Select Pressure Menu from the Test Menu. From the Press Menu screen
options select Let-by / Tightness.
2. Connect tube to positive pressure inlet but not pressure source.
3. Zero pressure, use the soft key marked P=0 .
4. Connect tube to pressure source to begin pressure test.
5. Press the soft key u to start test. Sprint displays the duration time on
the screen.
6. To stop test press the soft key ¢.
7. Press the soft key ü to pass test and proceed to Stabilisation test.
Let-by
14:44:27
Let-by
08-Feb-2006
Start P1
14:44:27
Let-by
08-Feb-2006
14:44:27
08-Feb-2006
0.0
Start P1
0.0
Start P1
0.0
Finish P2
0.0
Finish P2
0.0
Finish P2
0.0
Diff press
0.0
Diff press
0.0
Diff press
00:00
Duration
00:25
Duration
mBar
mBar
mBar
Duration
mBar
mBar
mBar
mBar
mBar
0.0
mBar
00:54
Passed?
P=0
Screen 1
Screen 2
Screen 3
You may fail the test by pressing the key û to indicate test has failed
(you can print the results).
Stabilisation Test
1. Press the soft key u to start test. Sprint displays the duration time on
the screen.
2. To stop test press the soft key ¢.
3. Press the soft key ü to proceed to Tightness test.
9
Sprint eVo1/eVo2/eVo3
Performing tests
Stabilisation
14:44:27
Stabilisation
08-Feb-2006
Start P3
14:44:27
08-Feb-2006
Stabilisation
14:44:27
08-Feb-2006
0.0
Start P3
0.0
Start P3
0.0
Finish P4
0.0
Finish P4
0.0
Finish P4
0.0
Diff press
0.0
Diff press
0.0
Diff press
00:00
Duration
00:25
Duration
mBar
mBar
mBar
Duration
mBar
mBar
mBar
mBar
mBar
0.0
mBar
00:54
Stabilised?
P=0
Screen 1
Screen 2
Screen 3
You may repeat the test by pressing the key û (you can print the
results).
Tightness Test
1. Press the soft key u to start test. Sprint displays the duration time on
the screen.
2. To stop test press the soft key ¢.
Screen 1
Screen 2
Screen 3
Screen 4
3. Press the soft key ü to pass test.
You may fail the test by pressing the key û to indicate test has failed.
At the end of the tightness test you can log or print the results. Press the ESC
key to return to Press Menu.
Press ESC again to return to Test Menu.
3. Pressure, Differential Pressure, Working Pressure and
Operating Pressure tests (eVo 2 and eVo3 only)
Before carrying out the test the pressure must be zeroed with the tubes connected to the instrument but not the pressure source. A thermocouple may be
connected to provide temperature readings in these tests.
10
Sprint eVo1/eVo2/eVo3
Performing tests
Tip: plug in the
1. Select Pressure Menu from the Test Menu. From
Thermocouple
the Press Menu screen options select Pressure, Diff
probe to view
Pressure. Working pressure or Operating Pressure.
temperature on
screen.
2. Connect tube(s) to pressure inlet(s).
3. Zero pressure, use the soft key marked P=0 .
Pressure
4. Connect tube(s) to pressure source(s) to
14:44:27
08-Feb-2006
begin pressure test.
5. There is an option to time this test, Sprint
Pressure
0.0
displays the duration time on the screen.
Duration
00:00
Press the soft key u to start test and timer.
Tip:
6. To stop timed test press the soft key ¢.
optional
7. To restart test press the soft key ¢tt.
timer
To re-zero the unit, press P=0 .
Press the ESC key to return to the Press Menu screen. To perform the other
pressure tests repeat steps 2 to 7 above.
Press ESC to return to the Test Menu.
mBar
P=0
4. Differential temperature test
Sprint can perform a differential temperature test with one or two thermocouple probes. To begin test, select Diff Temperature from the Test Menu. Check
the units displayed are the correct temperature scale. If no probes are connected Sprint will display ####.
Two probe test
Connect both probes to the k-type connectors. See label on back of unit for
FLOW and RETURN. Place probes in position. The screen will display the temperature of probe 1 and probe 2, and the differential
temperature.
Temperature
Use the soft keys to log or print the results.
14:44:27
08-Feb-2006
To end test press the ESC key.
Flow T1
####
C
Single probe test
Ret T2
####
C
When only a single probe is available, Sprint will display
an additional icon on the screen to represent the first and
second reading.
1. Place probe in position to make measurement T1.
to take a snapshot reading
2. Press the soft key icon
of T1. (Do not remove probe until this snapshot is
taken.)
T1
T2
Diff temp
####
C
T1
T2
T1
T2
Tip: this icon
will appear when
only one probe is
avaialble.
11
Sprint eVo1/eVo2/eVo3
Performing tests
3. Move the probe into position to take second reading T2. The screen
to show temperature reading T2 is being taken.
will display the icon
The screen will display the snapshot temperature, the live probe
temperature, and the differential temperature.
T1
T2
Use the soft keys to log or print the results.
To end test press the ESC key.
5. Room CO safety test
Ensure the unit has been zeroed in clean air before performing this test.
The instrument will proceed if already previously zeroed after switch on.
Note: For further details on performing Room CO Safety Tests, see Appendix I.
1.Select CO Room safety from the Test Menu. Sprint will display the
Appliance Menu. Select the appliance from the list.
The appliance will be displayed on the screen during the room CO
safety test.
2. Connect probe if required to the Sprint unit and place at the
recommended height. Refer to British Standard BS7967 if necessary.
3. The pump will switch on in readiness for test. Note: The sound of the
pump operating does not indicate the test has begun. Press the u soft key
to start test. During the test the screen will display the CO reading, peak
CO reading, duration of test and maximum allowed CO for the test.
The test will run for the required duration by appliance according to BS7967.
Sprint will emit an alarm if 30ppm (or 90ppm) threshold is exceeded. Sprint
is programmed with pass/fail criteria for this test. Refer to British Standard
BS7967 for further details on performing room CO safety tests.
When the minimum test period for an appliance has been reached (as defined
by BS 7967:2005 ) the ON/OFF/ENTER key can then be used to cycle through
CO Room
12:57:24
Time(Min)
CO
ppm
4
Peak CO
5
1
2
3
4
5
ppm
Allowed
ppm
Duration
of test
12
CO Room
1
12:57:24
23-Feb-2009
Water heater (flueless)
10
07.12
Duration
of test
CO Room
2
23-Feb-2009
CO(ppm)
1
2
5
4
4
07.12
12:57:24
Time(Min)
6
7
8
9
10
Duration
of test
3
23-Feb-2009
CO(ppm)
4
4
####
####
####
07.12
Sprint eVo1/eVo2/eVo3
Performing tests
the CO Room Safety screens. The test can be completed by pressing the ¢ soft
key to finish. Use the soft keys to log or print the results.
To stop test at any time, press the ESC key. Use the soft keys ü or û to select
'Quit test?'.
When printing Room CO safety test results Sprint will also provide data showing for how long the recommended level of CO was exceeded. This is expressed
as hh:mm > allowed.
6. Gas escape test (eVo2 and eVo3 only)
Note: Battery power will be used at a higher rate under this test.
Select Gas Escape Detection from the Test Menu. Connect the Gas escape probe
(GEP) to the USB connetor, the light will illuminate on the probe. Sprint will
display a warning if the probe is not connected.
Sprint will display 'sensor stabilising' for approximately 30 seconds.
When the sensor is stable the unit will ask 'in clean air?' before zeroing.
Press ü to zero the unit. Sprint will display a bar graph on the screen as gas
levels are monitored.
Place the probe in the area of inspection for several seconds before moving it
to other locations.
Sprint will emit continuous clicks like a Geiger counter. If higher gas levels are
detected the bar graph will increase in readings and the sounder will increase
in pitch.
Press ESC to quit the test.
As with many types of instrumentation the gas escape probe may be damaged
by impact. If the probe is dropped or suffers significant impact in another way
check its operation by first plugging it into the Sprint and confirming that it is
recognised by the analyser.
If this is successful select the gas escape mode and briefly apply a sample of gas,
for example from an unlit cooker ring. This will allow the user to verify whether
the sensor is working correctly or not. If the gas is not detected contact Telegan
or your local distributor to arrange for the probe to be serviced.
It is recommended good practice that the above check is carried out periodically
even if the user is not aware of the probe being exposed to impact.
If the Sprint detects a fault with the GEP it will display a warning advising that
the leak sensor is faulty and the test will be aborted.
13
Sprint eVo1/eVo2/eVo3
Operation
I. Operation
Before using the Sprint flue gas analyser on any heating appliance installation,
ensure you are familiar with the working practices defined in the BS7967 standard which details how to perform certain tests and safety issues to consider.
Auto zero setting
The Sprint unit will require auto zero before performing any of the following
tests: Flue Gas Analysis and Room CO Safety test. All other options will be available on start up. Ensure all tubes and probes are connected before entering an
option that requires auto zeroing and zero outside in clean air.
Pump
Sprint runs the internal pump during purge and when certain tests are entered
into; and during and after some tests. A rotating purge icon will appear on the
screen
when purging. The speed and therefore the sound emitted by the
pump, may vary depending on the test being conducted. To save on the battery
life, Sprint will turn off the pump when it is not required.
When the pump is running ensure the exit gas exhaust is not blocked and do
not breath in the exhaust gases.
1.1 Menu and operator button overview
Sprint provides a large graphic LCD with blue backlight. Navigation and functions are provided by three soft key buttons which change according to what
you are doing.
Use the soft keys to
navigate menus,
select, start and end
tests, change options
and select actions.;
such as print, log
or zero
Use the escape key
to exit menus, exit
tests and step back
through screens
14
ESC
ON/OFF/ENTER switch.
Use this key to finish
edits, save changes
and view next page of
multiple paged screens.
Sprint eVo1/eVo2/eVo3Operation
After your Sprint unit has been switched on the display will show the Test Menu
screen ready for use. To display the Main Menu press the ESC key.
To enter a menu item, use the soft keys below the é and ê screen icons to
scroll the menu list and press the soft key underneath the þ screen icon to
select. Some menus also have their own submenu.
Press the ESC key to exit a menu, press ESC twice to return to the Main Menu
from a submenu.
The soft keys control the function displayed above them on the screen. These
will change depending on the test or menu, or if the ESC key is pressed.
The menu structure is shown below:
Main menu
Press ESC to go
to Main menu
Flue Ga s
Analysis
Test Menu
Pressure
Diff
Menu
Temperature
evo2 and eVo3 only
Let-by / Tightness
Pressure
Diff Pressure
CO Room
Safety
Appliance
Gas Escape
Detection
Units
Pressure
Efficiency
Temperature
Settings
Auto off
timeout
Logs
Back light
Set date
& time
Select a log
Edit report
header
Key click
Print
Cal Due
Find a log by number
Flue CO
alarm
Supervisor
settings
Change
password
Delete all logs
How to display unit serial number, identity and software
version
Press the ESC key from the Main Menu
15
Setting up
Sprint eVo1/eVo2/eVo3
II. Setting Up
2.1 Configuration Options
Fuel Options:
Sprint displays the current fuel option on the screen. To change the fuel option
press the ESC key to enter the Main Menu and select Fuel Options. Use the soft
keys to scroll and select the desired fuel using ü. The new fuel option will be
displayed on the test screens.
Fuel options available are: natural gas, LPG, heavy oil, light oil, coal, wood,
wood pellet dry, coke, Biomass and Bagasse.
Note: NetHE only available for fuels natural gas, LPG, light oil and heavy oil.
From software versions i2.00, fuel types can be changed using Sprint PC Lite.
Units of measurement:
To change the units of pressure, temperature or efficiency press the ESC key to
enter the Main Menu and select Units of Measurement. Use the soft keys to
select the submenu and units for Pressure, Temperature or Efficiency.
Temperature units are: degrees Celsius (ºC) or degrees Fahrenheit (ºF).
Pressure units are: mBar, Pa, hPa, kPa, PSI, inWG, mmWG, inHG and mmHG.
Efficiency units are: Net, Gross or NetHE.
Note: the formulae and constants used for Gross and Net efficiency calculations are those specified in EN50379. In general it is Net efficiency which is
normally quoted. For modern condensing boilers the Net efficiency calculated
may exceed 100%. Sprint provides a condensing efficiency calculation via the
efficiency option NetHE. The result of this calculation takes into account the
recovered latent heat.
Analyser settings:
The Analyser settings menu allows you to alter the settings for the display, auto
off timeout, back light, key pad and Supervisor settings. To change any of these
settings press the ESC key to enter the Main Menu and select Analyser Settings.
Use the soft keys to select the submenus.
Auto off timeout
On the auto off timeout screen the number of minutes at which the unit will
automatically switch off is shown. Use the soft keys – and + to reduce or increase
the number of minutes. Press the þ key to accept the change or ESC to cancel.
16
Sprint eVo1/eVo2/eVo3
Setting up
To disable Auto off timeout press – key until 'disabled' is displayed.
Back light
The back light can be set to four options: off, dim, bright or controlled. Use the
soft keys to scroll and select the option desired. Press the þ key to accept the
change or ESC to cancel.
Back light options are:
Off: switches the backlight off
Dim: lowers the light intensity
Bright: increases the light intensity
Controlled: Sprint monitors the ambient light level and adjusts the light
intensity accordingly.
Reports (eVo3 only)
Allows selection of Printer (IR comms) or Bluetooth®. Use the soft keys to toggle
between these and the þ key to accept.
Key click
The Key click settings allows you to enable or disable the key pad from making
audible 'clicks' when pressed. Use the soft keys to select 'enabled' or 'disabled'
from the submenu. Press the þ key to accept the change or ESC to cancel.
Supervisor settings
Set date &time
On the Time & Date screen the current time and date are shown. Use the soft
keys – and + to alter the values of hours, minutes, day, month and year. Use the
è key to select each unit. Press the ESC key to accept the change.
Edit report header
Use the soft keys é and ê to select the report header text line one or two.
Press +
– key to edit text. The screen displays the character lists and highlights
the current list in use. Use the
and
keys to scroll through character values in each list and è to move on to the next letter in the header text. The
character lists are shown below.
Press ON/OFF/ENTER to delete characters to the right. Press è to move the cursor to the end of the text and þ to accept change and return to Edit report
header screen.
Print Calibration Due
Enable or disable printing of calibration due on reports.
CO Alarm
During flue gas analysis, a carbon monoxide alarm can be set to activate at
300ppm of CO or disabled.
17
Sprint eVo1/eVo2/eVo3
Setting up
Change password
Press the soft key +
– to edit password. Use the and keys to edit or create
a password as described above in Edit report header. Press þ to accept change
and return to Password screen. Press ON/OFF/ENTER to store the change.
When a password has been created, Sprint will display the Supervisor password screen on entering Supervisor settings.
! .. /
!"# $% & ' ( ) * + , - _ /
A .. Z
Uppercase alphabet
0 .. 9
0123456789
a .. z
Lowercase alphabet
: .. @
: ; < = > ? @
Space
Tip: It is a good idea to enter your name/company name and phone number to identify the
instrument and then set a password to prevent others changing these settings.
18
Sprint eVo1/eVo2/eVo3
Battery charging
III. Battery Charging
The Sprint has a lithium-ion rechargeable battery and will operate for up to
9 hours when fully charged, depending on the type of test used.
Low battery
When the battery is low, Sprint will display a low battery icon. If the battery gets
too low, then Sprint will give further warning before switching off.
Power Jack
Use of the charger as a power adaptor
The Sprint charger can be used to power the unit and will continue to charge
whilst operating the unit.
Automatic battery saver
The Sprint unit will automatically power down if left unused (unless Auto off
disabled). The auto off timeout can be set in Analyser settings. Sprint will warn
user when power down is imminent.
Charging the batteries
Warning:
Do not attempt to use any other charger with this unit except the one(s) supplied.
Failure to comply could invalidate the warranty and may result in permanent damage
to the unit.
1. Plug the charger into a mains socket.
2. Connect the charger to the Sprint using the power jack on the side of
the unit.
Switch on the power at the mains socket. The unit would normally be
left switched off for charging. The display will show the battery charging/
mains connection icon. When the battery is full both icons will flash.
19
Maintenance and calibration
Sprint eVo1/eVo2/eVo3
IV. Maintenance and calibration
General
To keep the display panel and operator buttons free from dirt build-up, regularly
wipe over your Sprint unit with a slightly damp cloth.
4.1 Unit
The Sprint should be calibrated once a year. Your Sprint unit will also display
the calibration due date at switch on and will warn you when calibration due
date is drawing near. If the calibration due date has passed, the Sprint unit will
display a message 'Calibration overdue'.
4.2 Water trap
The combined filter and watertrap is used in-line between the probe and the
unit. Before performing any tests, check that the filter is clean and there is no
water inside the trap.
Filter
Filter
The filter element should be changed if the filter has
become contaminated or dirty. If the filter has become
soaked with water, remove from trap and leave to dry
before reusing. Ensure 'O' ring remains in place.
To change the filter
Unscrew the filter-housing, remove the old filter and replace.
Warning: Filters
must be used at all
times. Failure to do
so may invalidate
the warranty.
Water
To empty the water trap, unscrew the filter-housing, remove middle disc and
o-ring, empty and dry with clean cloth. Replace central disc and o-ring as
shown in the diagram above.
20
Sprint eVo1/eVo2/eVo3Specification
V. Specification
Instrument
Operating temperature
range
Battery
-10°C to 50°C (14°F to 122°F)
Lithium-Ion. Life up to 9 hours dependant on test
used.
15% left warning.
Recharge time
Minimum 6 hours from flat.
Charger input voltage
230 V; 50 Hz AC
Standard Fuels
Natural gas, LPG, heavy oil, light oil, coal, wood,
wood pellet dry, coke, Biomass and Bagasse
Display
Blue back lit graphic LCD
Menu
Intuitive structure, tab selection on screen
Dimensions
92 x 176 x 59 (mm)
Function buttons/key pad 5 button keypad
Weight
528 g
Operating time:
9 hours (6 hours continuous pump)
Pump
Flow fail indication, SmartPurge
Enclosure
Integrated robust protective case.
Protective rubber boot with integral magnets
Standards
BS7927, EN50379, BS7967
Data Logging Reports
Up to 200 reports, depending on type
Probes
Standard Efficiency Flue
Insertion length
Maximum temperature
Construction
Probe
250 mm (9.9”) with adjustable depth gauge
800°C (1472°F)
Ergonomic pistol grip with stainless steel shaft,
in-built with thermocouple, in-line water trap/filter
K-type thermocouple
Accuracy ±1°C or ±0.3% of reading, whichever
is best
Protective shaft for thermocouple
Hose length
2000 mm
Gas Escape Sensor Probe
Gas Escape Sensor:
0-10,000ppm natural gas
21
Sprint eVo1/eVo2/eVo3
Specification
Gases
Range
Display
Resolution
Oxygen
0-25%
0.1%
± 0.2%
0.3% v/v
50 sec
30 sec
ü
Carbon
monoxide
0-5,000ppm
1ppm
≤20ppm;
±3ppm
45 sec
60 sec
ü
Carbon
dioxide
(Calculated)
0-25%
0.2% v/v
50 sec
30 sec
CO/CO2
ratio
0-10,000ppm
over range
0 to 0.9999
Accuracy
Detection
limit
Response
time
(t90)
Recovery Diagnostics
time
±3ppm
0.1%
±0.2% v/v
0.0001
Other Measurements
Measurement
Temperature
(selectable ºC or ºF)
Efficiency
XSAir
Range
-50º to 1100ºC (-58º to 2012ºF)
0-100% Net or Gross
0-120% Net High Efficiency
0-100%
Draught/Pressure Measurement
Let-by/Tightness Test
Dedicated test and report/print structure for
combined test to IGE/11/UP/1B
Pressure Scale
Range
Resolution
Accuracy
Equivalent scales
Communications
IR Port
USB
22
-200 mbar to +200 mbar
0.01 mbar
±0.5% of reading calibration at +50 mbar
(equivalent to ±0.5 mbar)
Pa, hPa, kPa, PSI, inWG, mmWG, inHG, mmHG
Sprint eVo1/eVo2/eVo3
Accessories and spare parts
VI. Accessories and spare parts
Accessory list
Telegan part number
Description
General Spares
CAS29023
PRB29002
PRB29000
PRB29011
TUB29000
CHG29001
FIL29001
INS29021/2/3
FIL99008
SPRINT EVO CARRY CASE
SPRINT V LEAK PROBE
SPRINT V MAIN PROBE INC WATER TRAP WITH FILTER
FREEVO MAIN PROBE INC WATER TRAP WITH FILTER
NEOPRENE TUBE – PRESSURE
SPRINT UK CHARGER
WATER TRAP INC FILTER
EVO INS CARD V1/V2/V3
FILTER ELEMENT
23
Logging
Sprint eVo1/eVo2/eVo3
VII. Logging
Sprint provides the option to log the results of tests. When a test is complete,
use the soft key < to log the results. The display will show the Create log screen
detailing the log number, log title, date and time. Press the accept key ü to
'Store log?'. The log details recorded are then displayed on the screen. Press the
þ key to continue or the soft keys to print or delete the log.
Recalling logs
To recall a log, use the ESC key to display the Main Menu. Logs can be recalled
either by viewing and selecting from the log list or by entering the log number.
The log list displays the log number, test title and date. Use the soft keys to
choose Select a log or Find a log by number.
When the log has been selected, Sprint will display the log on the screen. The
log data may be displayed on more than one screen. Press the  button to
scroll through screens. Use the soft key options to Print or delete the log. Use
the ESC key to exit the log list and Stored logs menu.
Printing logs using IR comms
Logs can be printed instantly from the Stored logs or directly after a log has
been recorded and the log is displayed on the screen. Press the Print soft key
to send the log file to the printer.
Deleting logs
Single logs can be deleted either via the Stored logs menu or directly after a
log has been recorded when the log is displayed on the screen. From the Stored
logs Menu choose Select a log or Find a log by number to find the log you wish
to delete. Delete the log using the soft icon.
All logs can be deleted from the Stored logs Menu. Scroll and select Delete all
logs and press the þ soft key. Press the yes ü key ' Are you sure?' to delete
all logs or press ESC to cancel. Deleting all logs resets next log number to one.
24
Sprint eVo1/eVo2/eVo3
Printing and Data Transfer
VIII. Printing and Data Transfer
Printing using IR comms
Ensure your selected printer is switched on with paper roll installed and ready
for use.
Ensure that the IR window on Sprint is aligned with the IR window on the
printer. The printer may be up to 1 m away from the Sprint unit. If a report can
be printed, a printer icon will appear on the screen. Press the soft key under the
icon for an instant print. Ensure printer is set to PC mode.
Printing can be aborted by a second press of the print soft key (when the print
icon is reversed) or by pressing ESC to exit the screen.
Data Transfer (eVo3 only)
Ensure Bluetooth has been selected from the Report menu. The Print button
will now display the Bluetooth symbol. Data can be transferred to a suitable
Bluetooth enabled PDA with the Sprint PDA application installed or PC running
Sprint PC Lite.
25
Sprint eVo1/eVo2/eVo3
Troubleshooting guide
IX. Troubleshooting guide
Sprint will provide on-screen messages which advise clear actions. Contact
Anton if unsure on how to proceed.
Symptom
Cause
Recommended User Action:
Instrument will not
turn on when on/off
button is pressed and
held for 2 seconds.
Battery flat
Connect charger and retry.
Battery may be flat. The instrument is designed to prevent
deep discharge occurring and
will turn it self off when battery
level gets too low.
Battery flattened
Instrument will not
beyond standard
turn on and charger
symbol is not displayed charging point.
when charger is connected.
Ensure charger is correct type.
If so, plug-in and leave connected. Check to see if charging symbol appears every 4-6
hours. If it does not and unit
does not switch on, return
both unit and charger for service.
Pump sometimes oper- Smart purge is operates at a faster rate.
ating. CO sensor is
recovering from exposure to gas.
Remove probe from flue during purge. Continue to use the
instrument as normal.
Pump flow fail alert
Filter/water trap or
sample line blocked.
Empty and clean filter/water
trap. Ensure sample line is free
from blockage.
Printer does not
respond or report contains odd characters.
Printer may be off,
faulty, out of range,
incorrectly set-up,
have a low battery or
not facing Sprint.
Ensure printer is charged up
and turned on, working, set-up
correctly with ‘PC’ protocol,
within physical range (usually
1m) and with the IR window
facing the IR window on Sprint.
(Other IR sources such as a PC
or sunlight may give odd character print-out)
26
Sprint eVo1/eVo2/eVo3
Troubleshooting guide
Symptom
Cause
Recommended User Action:
Auto-zero failed
Sensors exposed to
gas or faulty.
Switch off and on, ensuring
you zero in clean air and sensors are purged. If unit continues to fail auto-zero, return for
service.
Alarm activates in CO
Room safety test
CO is reading greater Dependant on safety procethan 30ppm or 90ppm dures.
(cooker).
Cannot remember the
supervisor password.
Forgotten
Supervisor password prevents
alteration of report header
which identifies owner of the
instrument. This provides security against theft. So there is no
method to remotely unlock the
instrument. Return to supplierfor reset
During auto zero the
gas reading does not
stabilise.
Recovering from high
gas exposure or gas
sensor faulty.
Ensure unit is purged and allow
sensor to recover or return for
service ASAP.
Sometimes negative
gas readings are displayed.
Previously zeroed with
gas present.
Turn off and on and repeat
auto zero in clean outside air,
allowing time for the CO sensor to recover and stabilise.
Sometimes “!>” or
“!<” is displayed in
place of a number.
Sensor is out of range. Contact support for advice.
Return for service if problem
persists or other failures are
observed.
Cal due date has
changed.
Time/date has been
amended.
Check current date and time
is correct. If the cal due date is
set to more than a year’s time
then return for service ASAP.
27
Sprint eVo1/eVo2/eVo3
Appendix I: Carbon Monoxide Room Safety Tests
Appendix I: Carbon Monoxide
Room Safety Tests
Sprint instruments are designed to assist heating engineers to work to the BS
7967 specification for carbon monoxide room safety testing. You should refer
to BS7967, which defines the requirements, details the methods as well as the
pass and fail criteria for various types of appliances.
CO Room Safety Test in the Sprint instruments is designed to measure the build
up of carbon monoxide levels in a room where a gas appliance is in use and
record those values each minute for the duration of the test. In addition at the
end of the test it assists the engineer (in an advisory capacity only) to determine
whether the test has passed or failed or if the results are invalid. In certain
circumstances, where the results are borderline or open to interpretation, the
instrument will ask the operator to decide if the test has passed or failed, and
will record the operators decision.
Please note: ultimately it is the responsibility of the operator to ensure that the test
is correctly performed to the BS 7967 specification. If the data does not support the
result or the operator suspects it is not reliable due to local conditions (such as carbon
monoxide level changes due to cigarette smoke or vehicle traffic) or incorrect, then
either the test should be repeated or the operator should seek expert advice.
CO Room Test Pass and Fail Test Specifications
Max Allowed CO:
Max Peak Duration
exceeding Max
Allowed CO:
CO Alarm Level:
Min Test Duration:
Max Test Duration:
Type C:
Type B:
Type A:
Type A:
Type A:
Room sealed
appliance
10 ppm
60 secs
Boiler
(open flue)
10 ppm
60 secs
Cooker
(flueless)
30 ppm
20 mins
Water heater
(flueless)
10 ppm
30 secs
Space heater
(flueless)
10 ppm
60 secs
30 ppm
15 mins
30 mins
30 ppm
15 mins
30 mins
90 ppm
20 mins
30 mins
30 ppm
5 mins
10 mins
30 ppm
30 mins
30 mins
Result Codes
The pass or fail result is displayed when the test completes and is printed on
the report as well as recorded in the log. If the test fails a code number is also
displayed, printed on the report and recorded in the log. This fail code identifies the way in which the test failed and can help identify the cause. Also when
28
Sprint eVo1/eVo2/eVo3
Appendix I: Carbon Monoxide Room Safety Tests
the test completes a short text message associated to this code is displayed in
a pop-up prompt dialog screen, to explain the reason for failure.
The result codes and associated prompt dialogue messages are as follows:
RESULT & CODE
POP-UP PROMPT ON-SCREEN
“PASSED"
“PASSED (2)"
ANY “FAILED”
“FAILED (1)”
“FAILED (2)”
“FAILED (3)”
“FAILED (4)”
None
None
“Warning - CO Room Safety test failed.”
CO levels did not fall or unstable.
CO unacceptably high (for too long).
CO dangerously high.
Unacceptable or incomplete.
“Press ESC key to continue.”
Pass Cases
Normal Acceptable Peak of CO
Normally for a test to pass, the CO levels must peak without exceeding the
maximum allowed CO level and then fall (by at least 1 ppm) below the peak
value before the end of the test. Note that it is not necessary for the CO level
to reach or be close to zero at the end of the test, so long as it remains below
the maximum allowed CO level.
The result code is: PASSED
Very Low Levels of CO
If the CO levels remain below 3 ppm (ie: close to clean air or background noise
levels) for the duration of the test, then the test is considered to have passed.
The result code is: PASSED
Failure Cases
Excessive Levels of CO
If the CO level exceeds the CO alarm level then the test is considered to have
failed and should be immediately aborted. The CO alarm may be triggered at
any stage (before, during and after completion of the test) and continues to
annunciate until the CO level returns to a safe level.
The result code is: FAILED (3)
29
Appendix I: Carbon Monoxide Room Safety Tests
Sprint eVo1/eVo2/eVo3
NB: The alarm should prompt the operator to take appropriate action according to
BS7967 and safety procedures. The sensors in the instrument should be purged with
clean air and allowed to recover.
Unacceptable Levels of CO
A peak duration timer records whenever the CO level exceeds the maximum
allowed CO level during the test. If the total peak duration time exceeds the
max peak duration allowed then the test fails due to unacceptable levels of CO.
The result code is: FAILED (2)
Operator Pass/Fail Cases
The following results are considered to be operator determined whether the
test passes or fails:
Acceptable Levels of CO with no Peak
For some appliances CO levels may rise to a value beneath the specified limit
and stabilise rather than fall. In this case it is up to the operator to determine
PASS or FAIL.
If operator chooses to fail the test, the result code is: FAILED (4) otherwise the
result code is: PASSED (2)
CO Level Exceeds Max Allowable Level for a Short Duration
For some appliances (eg: cooker) a peak exceeding the maximum allowed CO
level may be acceptable, provided the CO level falls back below this level within
the max peak duration time. It is up to the operator to determine if the result
is acceptable or not.
If the operator chooses to fail the test, the result code is: FAILED (4)
Otherwise the result code is: PASSED (2)
Other Cases
When a test is not performed correctly, the results are inconclusive or there is
insufficient or unreliable data Sprint attempts to interpret the readings detected
and fail the test. It is possible for the results of a test to appear to be valid when
it was performed incorrectly or the data collected was unreliable in some way.
Please refer to BS7967 and ensure tests are carried out correctly. Sprint V tries
to reject incorrectly taken test readings but should not be relied on to instruct
engineers on correct working practice.
30
Sprint eVo1/eVo2/eVo3
Appendix I: Carbon Monoxide Room Safety Tests
Multiple Peaks of CO
The overall peak CO reading recorded will be the latest peak CO reading that
was higher than any previous. The peak duration recorded will be the total
time the CO readings exceeded the maximum allowed CO level. The pass / fail
criteria are applied as before based on this information. Sprint does not expect
to record multiple peaks of CO where the reading goes up and down more
than once.
Unstable or Rising Levels of CO
If there is a significant build up of CO levels at the end of the test and levels are
still rising the test will fail.
The result code is: FAILED (1)
CO Level Exceeds Max Allowable Level and Peaks for Unknown
Duration
If the CO level exceeds the maximum allowed too close to the end of the test
then the test is failed. This can occur if there is a build up of CO levels towards
the end of the test or the appliance fires up late on in the test or the test is
stopped too early. The test should be repeated if this occurs.
The result code is: FAILED (2)
CO Level Does Not Start Close to Zero
It is important that the instrument is zeroed in clean air at switch-on. Failure to
do so will invalidate the test result. BS 7967-2 also requires that the room is
well ventilated prior to starting the test. However it is possible that there may
be a residual background CO level (eg: due to traffic fumes). The instrument
will display a pop-up box prior to starting the test if the CO level is more than
3 ppm.1
31
Technical helpline: Tel: +44 (0)870 428 0075
Manufactured by:
Telegan Gas Monitoring
A division of Crowcon Detection
Instruments Ltd
2 Blacklands Way,
Abingdon Business Park
Abingdon
Oxfordshire OX14 1DY
United Kingdom
Tel: +44 (0)1235 557700
Fax:+44 (0)1235 557749
Email: [email protected]
Web site: www.telegangas.co.uk
Exclusive UK Distributor:
Anton Industrial Services Ltd
Unit 6 Greenhill House
26 Greenhill Crescent
Watford Business Park
Watford.
WD18 8JA
United Kingdom
Tel: +44 (0)870 428 0075
Fax:+44 (0)870 428 0076
E-mail:[email protected]
web: www.anton-group.com