Download Technical Reference Manual LFXG-D
Transcript
LFXG-D setup Process standardization configuration Standardization adjusts the system by resetting one point of the calibration curve to an independently measured or known level. The frequency of standardization depends on several factors, including desired accuracy of the reading. Set up the following sub-parameters for standardization: STANDARDIZE_POINT STANDARDIZE_TYPE Standardize point Process value recorded for the last standardization. Standardize type Process standardize type determines how you enter the actual process value of a standardize sample. If this is set as Use Lab sample value, the software screens prompts entry of the sample value during a standardize. If this is set as Use Default value, the software always uses the Default standardize level as the sample value. Standardize default value Default standardize is the default level value in engineering units that you use in the standardization procedure. At standardization, enter the actual level of the process material to override this default. Standardize interval Standardize interval is the interval in days between standardize alarms. The level transmitter alarms to indicate that a standardize procedure is due if the diagnostic alarm, Standardize due, is set. 38 LFXG-D Technical Reference Manual