Appendix D Mobile Phase Absorbance This appendix provides a list of the absorbances at several wavelengths for commonly used mobile phases. Choose your mobile phase carefully to reduce baseline noise. The best mobile phase for your application is one that is transparent at the chosen detection wavelengths. Such a mobile phase ensures that any absorbance is due only to the sample. Absorbance by the mobile phase also reduces the linear dynamic range of the detector by the amount of absorbance that is autozeroed out. Wavelength, pH and concentration of the mobile phase will affect its absorbance. Examples of several mobile phases are provided below. Table D-1 Mobile Phase Absorbance Measured Against Air or Water Absorbance at Specified Wavelength (nm) 200 205 210 215 220 230 240 250 260 280 Acetonitrile 0.05 0.03 0.02 0.01 0.01 <0.01 — — — — Methanol (not degassed) 2.06 1.00 0.53 0.37 0.24 0.11 0.05 0.02 <0.01 — Methanol (degassed) 1.91 0.76 0.35 0.21 0.15 0.06 0.02 <0.01 — — Isopropanol 1.80 0.68 0.34 0.24 0.19 0.08 0.04 0.03 0.02 0.02 Unstablized Tetrahydrofuran (THF, fresh) 2.44 2.57 2.31 1.80 1.54 0.94 0.42 0.21 0.09 0.05 Unstablized Tetrahydrofuran (THF, old) >2.5 >2.5 >2.5 >2.5 >2.5 >2.5 >2.5 >2.5 2.5 1.45 Solvents D 70 Appendix D