Download Data sheet - Instrumart
Transcript
4 Quick Start Setting the Temperature Press “SET” to select this memory and access the set-point value. S Access set-point value C 25.00 Current value of set-point 1, 25.00°C Press “UP” or “DOWN” to change the set-point value. U Increment display C 30.00 New set-point value Press SET to accept the new value and display the vernier value. The bath begins heating or cooling to the new set-point. S Store new set-point, access vernier 0.00000 Current vernier value Press “EXIT” and the bath temperature will be displayed again. E Return to the temperature display 24.73 C Bath temperature display The bath heats or cools until it reaches the new set-point temperature. Turn off the cooling to reach and control at higher temperatures. When setting the set-point temperature be careful not to exceed the temperature limit of the bath fluid. The over-temperature cutout should be correctly set for added safety. See Section 9.10, Cutout. To obtain optimum control stability adjust the proportional band as discussed in Section 9.9, Proportional Band. 15