COMMAND: XCo
XCos SYNTAX:
aXCo!
aXCos!
Where:
a = address of probe
s = signed slope
o = signed offset
Rev. 07/13
(Keller Digilevel mode only)
(Keller Digilevel mode only)
IMMEDIATE RESPONSE:a0012
RESPONSE TO D0:
aso
Where:
a = address of probe
o = sensor offset
s = sensor slope
FUNCTION:
Set sensor offset and slope in calibration lab environment.
EMULATION MODES:
Keller Digilevel
DESCRIPTION:
The probe’s sensor can be calibrated to account for accuracy and offset drift. Off
set, as used by this command, is not field or stage offset. Offset here is used to zero
out depth/pressure measurements. Slope is used to correct pressure readings at
some pressure, usually near the maximum.
Previous examples used the sensor depth/pressure reading as part of an equation
(see C, C1, M, and M1 for examples). Part of the equation included user slope and
offset. The offset and slope set by this command used before any user or field
slope and offsets. It is used as follows:
Sensor_reading = Raw_sensor_reading * XC_slope + XC_offset.
Calibration process should start by resetting the probe. This can be done by issuing
an aXWCO+1! command. After this command is issued, the probe’s address will reset to 0 (zero). User slope, offset, and units of measurement are also reset.
After issuing aXWCO+1!, the probe must be set up again.
Slope and offset may have been set by Keller at time of manufacture using this
command. Slope and offset set by XC command should be reset to 0.000 (offset)
and 1.000(slope). The following command should be issued.
0XC+0+1! - This command is issued multiple times until satisfactory results are
obtained.
Next, set the offset. Obtain the current reading (should be at atmosphere) using
the M! command. Take the current reading, negate it, and use it as the offset for
the next XC command. Issue the M! command and get the reading. Ideally, current
reading should be 0. If not, adjust offset as necessary.