Download Genesis -User Manual
Transcript
Genesis Operator’s Manual INDICATIONS • The gas flag will show “WEAK” at the conclusion of the calibration period. RESPONDING TO A WEAK SENSOR 1. If you are manually calibrating the Genesis, verify that you are correctly adjusting the gas concentration to match the concentration of the calibration gas. If using the Auto Cal method, verify that the gas concentration in the calibration gas cylinder matches the Auto Cal Gas Value. See Chapter 4, Setting Auto Cal Gas Values. NOTE A weak sensor is still functioning properly and can continue to be used. However, it is near the end of its useful life, and should be replaced as soon as possible. See Chapter 6, Replacing/Adding a Sensor, for replacement instructions. 44 71-0132