Download Standard Operation Procedure - Ru

Transcript
Standard Operation Procedure for detection of Color in water;
Procedure for Operation and Calibration for Color Determination of Water by Visual Comparison
Method
PURPOSE:
 To establish a procedure for examination of true (not apparent) color of water.
SCOPE:
 This procedure is applicable for determination of true color of water only.
RESPONSIBILITY:
 WQL Analyst: Operation and Calibration.

WQL Manager: Receiving sample & log book entry and Implementation of this SOP.

WQ Consultant: QC/QA.
DEFINITIONS:
 Apparent color is the color of the whole water sample, and consists of color from both
dissolved and suspended components.

True color is measured after filtering the water sample to remove all suspended material. The
term color is used here to mean true color.
HEALTH AND SAFETY:
 Personal protective equipment including safety glasses and lab coats should be used.
CAUTIONS:
Pretreatment for Turbidity Removal:
To determine color, turbidity must be removed before analysis. Methods for removing turbidity
without removing color vary. Filtration yields results that are reproducible from day to day and
among laboratories. However, some filtration procedures also may remove some true color.
Centrifugation avoids interaction of color with filter materials, but results vary with the sample
nature and size and speed of the centrifuge. When sample dilution is necessary, whether it
precedes or follows turbidity removal, it can alter the measured color. Acceptable pretreatment
procedures are included with each method. State the pretreatment method when reporting
results. [1].
Standard Operation Procedure Manual for Analysis of Drinking water quality
Page 23 of 64