Download 8.2 Optical unit
Transcript
11002644_AA.book Page 168 Monday, February 21, 2005 4:45 PM 11 Reference information 3 Select printing to a recorder or a computer and press OK. (UV) 4 The current setting for channel 1 is displayed. Press OK to access the setting menu. Set Rec Out 1 (UV) UV pH Cond %B Tmp Pr 5 Select the parameter to be printed on channel 1 and press OK. 6 Repeat steps 4 and 5 for channels 2 and 3. (no) 7 If the UV-curve is selected, the current setting for auto-scaling of the UVcurve is shown. Press OK to access the setting menu. (no) no 8 Select the desired setting and press OK. Print out to? Recorder Computer Set Rec Out 1 Autoscale UV Autoscale UV yes The three selected curves are now printed. Setting analog outputs See section 8.6.5 Setting analog outputs. Calibrating the flow rate See section 9.20.1 Calibrating the pump. Calibrating the pressure offset See section 9.20.2 Calibrating the pressure sensor. Set up show pH Normally, the pH is displayed in the running window. If not required, the pH display can be set to off. Set Parameters 1 From the main menu, select sub menu Set Parameters and press OK. Setup and calibration 2 Select sub menu Setup and calibration. Press OK. Setup pH 3 Select sub menu Setup pH and press OK. Set Show pH (on) 4 Select sub menu Set Show pH. The current setting for showing pH is displayed. If on is shown, current pH is displayed in the running display. If off is shown, no pH is displayed in the running display. Press OK to change the setting. 5 Select the desired setting and press OK. Set Show pH (off) 168 on off ÄKTAprime plus User Manual 11-0026-44 Edition AA