Download Instruction Manual Operator`s

Transcript
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
13
Breathing Pressure Monitoring
Breathing
Pressure
Monitor
Window
The Breathing Pressure Monitor window, shown in Figure 13-3, contains
breathing pressure measurements expressed in units of cm H2O.
Figure 13-3. Breathing Pressure Monitor Window
HI:APNEA - VOL
ME:MIN VOL LOW
LO:THRESHOLD LO
TID VOL
BPM
0.71 11
50
OP00310
OXYGEN
MIN VOL
33
7.8
100
16
1.0
PEAK BREATHING
PRESSURE
PEAK
49
12
8
MEAN
PEEP
12
MEAN BREATHING
PRESSURE
PEEP (POSITIVE END
EXPIRATORY PRESSURE)
•
PEAK Breathing Pressure—The highest instantaneous pressure
value for each breath
•
MEAN Breathing Pressure—The average of all the instantaneous
pressure values recorded during each breath
•
PEEP (Positive End Expiratory Pressure)—The breathing
pressure at the end of exhalation
During apneic conditions, the pressure monitor displays numeric
information as long as it detects a peak pressure at least 10 cm H2O
greater than PEEP pressure. When this pressure difference drops below
10 cm H2O, the numeric information remains 1 minute longer and then
disappears.
13-6
Narkomed MRI Operator’s Manual
Part Number: 4113877-003 Rev: F