Download USER MANUAL
Transcript
Troubleshooting If the Power Vac Aspirator fails to function, consult the Troubleshooting Guide below. There are no serviceable parts inside. DO NOT open case. If problem cannot be solved, consult your Provider. Problem Probable Cause Remedy The Aspirator will not operate 1. On/Off switch is in the “OFF” position 2. Power Cord is not connected properly 1. Push On/Off switch to the “ON” position 2. Check plug at wall receptacle and AC Inlet on the back of Aspirator 3. Shut the Aspirator “OFF”. Wait 15 minutes, check that air filters are clean and clear 4. Replace Fuse 3. Thermal limit exceeded 4. Blown Fuse With Compressor running / no vacuum at Patient Connection Tubing 1. Control Knob completely turned counterclockwise 2. Hydrophobic Filter assembly not connected properly 3. Suction Canister not connected or assembled properly 4. Defective Suction Canister 1. Increase vacuum with Control Knob No vacuum on gauge/ vacuum at Patient Connection Tubing 1. Defective vacuum gauge 1. Contact your Provider Vacuum on gauge/ no vacuum at Patient Suction Tubing 1. Clogged Hydrophobic Filter assembly 2. Canister overfilled 1. Replace Hydrophobic Filter assembly 2. Empty Canister Unable to adjust vacuum 1. Defective Vacuum control 1. Contact your Provider 2. Confirm and tighten connections 3. Confirm and tighten connection/assembly 4. Replace Suction Canister REPLACEMENT PARTS LIST Description Part # Patient Suction Tubing Inline Hydrophobic Filter & Tube Assembly Inline Hydrophobic Filter Suction Canister - 2000 cc (with lid) Small Filter Large Filter Fuse 1955 502690 1690 502687 502391 502471 503240 ASPIRATOR