Download DA100 Data Acquisition Unit User`s Manual
Transcript
4.5 Specifications of Strain Input Module Style Number: S5 Model Code, Number of Input Channels, Terminal Configuration, Shortest Measurement Interval and Values of Built-in Resistors Model Code Number of Input Channels Terminal Configuration Shortest Measurement Interval Built-in Resistor DU500-12 DU500-13 DU500-14 10* 10* 10* Clamp Clamp NDIS** 0.5 s 0.5 s 0.5 s 120 Ω 350 Ω External *: Requires the space of two slots. **: Recommended by the Japanese Nondestructive Inspection Association Method of Input Non-balanced floating input, with isolation between channels (separated channels) Resolution of A/D Conversion Equivalent to reading on 20000 full scale Integral Time of A/D Conversion Manual or automatic selection between 20 ms (50 Hz), 16.7 ms (60 Hz) and 100 ms (10 Hz) However, AUTO does not function if the instrument is the stand-alone DC power supply model, or the subunit (DS400/DS600) of the DC power supply model. (Selecting “AUTO” will set the A/D integration time to 20 ms (50 Hz)). Types of Input Signals from strain gauge or strain-gauge sensor Connection Methods Single-gauge, opposed-side two-gauge, adjacent-side two-gauge or four-gauge configuration (Configured with built-in DIP switches for DU500-12 and DU500-13) Applicable Gauge Resistance 100 to 1,000 Ω The DU500-12 and DU500-13 modules contain 120Ω and 350- Ω resistors, respectively. Bridge Voltage Fixed to 2 V DC ( ±5% accuracy with a correction function) Gauges 2.00 (with scaling capability) Balancing Automatic Range of Balancing Equivalent to ±10,000 µε (for single-gauge method) Measuring Ranges and Accuracies Reference operation conditions temperature: 23 ±2°C; humidity: 55 ±10% RH; supply voltage 90 to 250 VAC (AC power supply) /10 to 32 VDC (DC power supply); supply frequency: 50/60 ±1% (AC power supply); warm-up time: at least 30 minutes; when operating, the system must not adversely affect the operation of other equipment. Gauge Method Resolution Measuring Range Accuracy Single-gauge –2000 to 2000 µε –20000 to 20000 µε –200000 to 200000 µε 0.5% of Range 0.3% of Range 0.3% of Range 0.1 µε 1 µε 10 µε Two-gauge –1000 to 1000 µε 10000 to 10000 µε –100000 to 100000 µε 0.5% of Range 0.3% of Range 0.3% of Range 0.1 µε 1 µε 10 µε –500 to 500 µε 0.5% of Range 0.1 µε –5000 to 5000 µε 0.3% of Range 1 µε –50000 to 50000 µε 0.3% of Range 10 µε The Accuracies and Resolutions of the DU500-12 and DU500-13 models, do not include the influence of the internal-and contact-resistance of the jumper setup switch. Four-gauge 4-18 IM DA100-01E