Configuration and calibration
Figure 5.4.3.5.3.1
If you select the CANCEL option, the device returns to the previous screen without making any
change. However, if you select CLEAR CORNER ADJUST and press
, the device will delete any
corner adjustment previously made (it will return the span values of all the cells to 1.000000).
5.4.3.6 DIAG. DCELLS
A menu with several diagnostic functions for the load cells. Read-only
.
5.4.3.6.1 COUNTS
The first screen shown upon accessing the DIAG.DCELLS menu is one that shows the counts
of each cell followed by the percentage that these represents from the scale total; to the right, the total
points and gross weight. The screen format is the following:
Figure 5.4.3.6.1.1
The load cell that provides a greater signal is highlighted.
To show the screen with the next diagnostic function, press
one, press
. Press
to return to the previous menu. Press
menu.
and to go to the previous
to completely exit the SETUP
5.4.3.6.2 STATUS
This screen shows the status of each cell. If everything is correct, “OK” appears. If there are
one or more errors, the following texts are shown highlighted with an inverted background:
Text
“SN”
Error
The serial number does not coincide with the NVM value (adjustment
value).
The span does not coincide with the NVM value (adjustment value).
“GA”
The cell capacity does not coincide with the NVM value (adjustment value).
“CA”
The cell NOM does not coincide with the NVM value (adjustment value).
“NO”
The cell status indicates some kind of error.
“ST”
"NOT CONNECTED" There is no communication with load cell
If there is more than one error, they are shown separated by hyphens. For example, a serial
number and span error: “SN-GA”.
The screen format is the following:
5-29