Download Sartorius Cubis Series - Laboratory
Transcript
Statistics Statistics Purpose: Saving and statistical analysis of weight values and other calculated values. The values are generated as results: – Transaction counter – Sum of all values – Average value – Standard deviation – Lowest value (minimum) – Highest value (maximum) – Difference between maximum and minimum Options: You can save a maximum of 65535 components. Combination options: D Density determination, calculation, averaging, weighing in percent, parts counting, timer-controlled functions, checkweighing, mass unit conversion, second tare memory, individual identifiers (SQmin function, DKD measurement uncertainty) Configure Task: Statistics * = factory setting t Define how the weight value will be applied. *Manually: The weight value is applied as soon as the corresponding button is touched. Automatically: first value at stability: The weight value is applied automatically as soon as the balance has stabilized after filling and the value exceeds the minimum load. Automatically: last value at stability: The last weight value with stability is applied automatically, as soon as the balance is unloaded and the value falls short of the minimum load. If stability was reached repeatedly between measurements, the last stable weight value will be stored. You must now select the minimum load to be applied: (none, 10, 20, ...*100 ... 1000 digits). Weighing Mode with Automatic Storage: t Select the weighing mode to be used for the weight values. Initial weighing: Initial weighing of samples in a container. Final readout: Final weighing of samples in a container. 84 Cubis MSA User Manual