Download HVM100 Manual - PC Environmental Ltd
Transcript
nels) based on the sensitivity of the accelerometer. The values in the table assume that a calibrator that vibrates 1g rms and 159.15 Hz, will be used. Gain setting for various accelerometer sensitivities Gain 1 mV/g 10 mV/g 100 mV/g 60 dB 40 dB 20 dB Step 4 Connect a DC rms voltmeter to the DC output. Step 5 Turn on the calibrator and start the HVM100 running. The HVM100 should display an Arms level of 1 g. Step 6 Note the voltage reading on the voltmeter. The voltage corresponds to 1 g rms of acceleration. The voltage will typically be about 600 mV for a 1 g rms input signal. Step 7 Stop and Reset the HVM100. Reduce the calibrator’s output level to 0.1 g rms. If this isn’t possible, change the HVM100’s weighting to Wh (Hand-Arm mode). The Wh weighting will attenuate the calibrator’s signal to 0.1 g rms (an attenuation of 20.0 dB). Step 8 Turn on the calibrator and start the HVM100 running. The HVM100 should display an Arms level of 0.1 g. Step 9 Note the voltage reading on the voltmeter. The voltage corresponds to 0.1 g rms of acceleration. The voltage will typically be about 500 mV for a 0.1 g rms input signal. Step 10 The DC output uses a logarithmic scale. The formula for the sensitivity is: (voltage at 1g) - (voltage at 0.1g) DC output sensitivity = __________________________ 20 dB E-2 DC Output Calibration HVM100 Manual