Download - Frank`s Hospital Workshop

Transcript
11. Software
Call the Service Mode (see Chapiter 7). The display will
then read:
SET=Protocol
"Analyzer is in the Service
Mode"
Inquire the current software version, date, time and error
number via a return from service PC to the analyzer.
Miditron® Junior II will answer to the terminal program:
VER y xx.xx
Program-No. Y=1 (International)
Program-N0. Y=2 (USA)
Software-Version X.X
DAT tt.mm.yy hh:mm Date and time
ENR xxx
Error number ( 0 = no error)
Table: tab-9e
Enter the keyword "PW3308" via the terminal program. The
display reads:
SW update ?
"Software update can be
selected"
Press any key to cause the analyzer to change to the
"Software-Update" mode. Press <Set> until the following is
displayed:
Source: PC
"Software update via upload can
be selected"
erasing Programm
"Old software is erased"
Break: no Update
"Software update was aborted"
Send 1. File
"The first file is sent"
loading Program
Press <Enter> to start the software update. The display
reads:
The existing software is erased.
The software update can be aborted by pressing <Enter>.
The display reads:
If the existing software has not been deleted, press <Set>
to cause the analyzer to change to the "Source: PC" display.
The three hex files juup.h00, juup.h01 and juup.h02 are
transferred one after the other via upload. The display will
always read the number of the file that is being transferred
at a time.
"The file is being sent"
Send 2. File
"The second file is sent"
The software update can be aborted with <Set> or Power
Off/On. The display will then read again "SW update ?".
Service Manual Miditron® Junior II / ID 1997491 / MJ / 1.0 - June 1998 / Page 94