Download VIO 300 D

Transcript
8 • Bipolar Standard Modes
Diagram
BIPOLAR SOFT COAG ++
250
Effect8
200
Power (W)
Effect7
Effect6
150
Effect5
Effect4
100
Effect3
Effect2
50
Effect1
0
10
100
1000
Resistance (Ohm)
Fig. 8-10
Properties
AUTO START
In the Select activation type window, you can select an AUTO START function for
BIPOLAR FORCED COAG. When the instrument touches tissue, coagulation
starts automatically after a specified period of time.
Areas of use
All bipolar coagulation procedures in which you want to coagulate vessels fast and
effectively or want to replace monopolar forceps coagulation.
Difference from BIPOLAR SOFT
COAG
Suitable electrodes
Technical data
100 / 170
Fast bipolar coagulation.
Faster bipolar coagulation. Carbonization of the tissue cannot be precluded.
Bipolar instruments, e.g. bipolar forceps, bipolar hook electrodes.
HF voltage waveform
pulse-modulated sinusoidal alternating voltage
Rated frequency
350 kHz (at RL= 500 ohms) ± 10%
Crest factor
3.8 (at RL = 500 ohms)
Rated load resistor
200 ohms
Max. HF peak voltage
560 Vp
Number of effects
2
Constancy of effects
automatic control of HF peak voltage
80113-334
10/ 2009
BIPOLAR FORCED COAG