Download 3 - Agilent Technologies
Transcript
8 Abundance <- [M + NH ] + 4 100000 80000 60000 40000 20000 m/z 603 400 987 639 600 800 1000 Figure 74 Mass spectrum of the fatty acid triolein (C18:1, [cis]-9) molecular weight = 884.781 molecular formula = C57H104O6 Refractive index detectors APCI requires some compound volatility and is less suitable for highly thermally labile compounds. Figure 74 shows a typical triglyceride mass spectrum. Both the degree of unsaturation and the length of the fatty acid side chains can be determined from the [M + NH4] + ion, which corresponds to mass M + 18. In-source CID experiments also can be helpful in determining the fatty acid composition of chromatographic peaks. Full-scan methods allow easy identification at the low nanogram level. If more precise quantitation is required, selected ion mode (SIM) can be used to obtain detection limits at the low picogram level. ✔ Polar and semipolar compounds up to 150,000 daltons can be analyzed. Highly sensitive. Strong molecular ions. Fragments, depending on in-source CID parameters. ✘ Data analysis for complex heterogeneous mixtures of multiply charged analytes is not straightforward. Matrix can interfere with the ionization process. Refractive index (RI) detection is based on the difference in RI between the solution in the sample cell and the pure mobile-phase solution in the reference cell. Because the composition of the eluents must remain fixed throughout the analysis, this detector is not suitable for gradient analysis. Four main types of RI detectors are available: deflection according to Snell’s law, reflection according to Fresnel’s law, interference, and Christiansen effect. The first, which uses the dual-cell design, is by far the most popular. However, the nearly designed Agilent 1100 Series refractive index detector allows detection limits to the low ng range. Because RI detectors lack sensitivity and exhibit a tendency to drift owing to temperature changes, they are used primarily in the analysis of carbohydrates and nonaromatic acids. ✔ Universal detector. 104 ✘ Low sensitivity, no gradient operation.