Download 3 - Agilent Technologies
Transcript
Abundance 6 4 5 3 12 0 7 8 1 2 3 4 5 6 7 8 Aldicarb sulfoxide Aldicarb sulfone Methomyl 3-hydroxycarbofuran Aldicarb Carbofuran Carbaryl Methiocarb 10 Time [min] 20 Figure 72 Carbamate analysis and, as they shrink, analyte ions are desorbed. The ions are transported to the mass analyzer through a series of vacuum stages and ion-focusing elements. Electrospray ionization can produce multiply charged ions of macromolecular analytes such as proteins and peptides. Because mass analyzers separate ions based on mass-tocharge ratio (m/z), lower-cost mass spectrometers with mass ranges of several thousand m/z can be used to analyze compounds in excess of 150,000 daltons. The primary use of electrospray has been the analysis of compounds of higher molecular weight. However, this technique also has been applied successfully to small polar molecules. Fig. 72 shows a separation of carbamate pesticides using electrospray. HPLC inlet Nebulizer Skimmers Octopole Capillary + + + + + Corona needle Fragmentation zone (CID) Lenses Quadrupole Figure 73 APCI LC/MS interface APCI also can be used to analyze moderate polarity analytes. As in electrospray, APCI ionization occurs at atmospheric pressure via a chemical ionization process (see figure73). 103