B40/B20 Patient Monitor
Invasive blood pressure
Measurement range
-40 to 320 mmHg (-5.3 to 42.7 kPa)
Measurement accuracy
±5% or ±2 mmHg, whichever is greater
Frequency response
4 to 22 Hz
Transducer sensitivity
5 µV/V/mmHg
Pulse rate
Range
30 to 250 bpm
Accuracy
±5% or ±5 bpm, whichever is greater
Display resolution
1 bpm
Zero adjustment range
± 150 mmHg
Temperature
Measurement units
° Fahrenheit (F)
° Celsius (C)
Measurement range
10 to 45°C (50 to 113°F)
Measurement accuracy
± 0.1°C without temperature sensor
Display resolution
± 0.1°C at 25 to 45 °C with reusable probes
Probe types supported
Use only GE Healthcare recommend temperature
YSI probes.
Temperature self-check
At start-up and then every 10 minutes
Probe type time response
Reusable skin temperature 3 s
probe:
2-32
Reusable adult central
temperature probe:
6s
Reusable pediatric central
temperature probe:
4s
Disposable skin
temperature probe:
3 to 6 s
Disposable central
temperature probe, 12F:
5 to 8 s
Disposable central
temperature probe, 9F:
5 to 8 s