Push-To-SetTM Vacuum Regulators
The Ohio Medical® line of Push-To-SetTM (PTS) vacuum regulators are specially designed for a variety of clinical applications including airway,
nasogastric, surgical suctioning and C.A.S.S. procedures as well as chest drainage. Ohio Medical has been providing suction products to
healthcare facilities worldwide for 50 years. Our vacuum regulators continue to be the preferred choice for maximum performance, high
quality and minimum maintenance. The vacuum regulators below are used in product configurations on pages 4-13.
All Push-To-SetTM Vacuum Regulators feature:
•
Patented Push-To-SetTM technology for accuracy and
•
Safe, moderate to high flow rates
safety
•
Glow-in-the-dark gauge needle and increments (Analog)
•
Quick-To-Max - 1 to 2 turns to full wall vacuum for
•
Dependable, time-tested gauges and controls
emergency resuscitation
•
Low maintenance, easy service
•
Lightweight, compact design
•
Comprehensive ten year parts and labor warranty (Regulator only)
•
Impact-resistant, corrosion-free, rugged construction
•
Now available with a Digital Gauge (+/- 1% Gauge Accuracy)
•
Easy-to-read, color-coded gauge ranges
•
“Battery Low” Indicator on the digital gauge
Three Mode
Continuous
(PTS-CVR)
Adult
Three Mode
High
(PTS-CVR)
Adult
Two Mode
Continuous
(PTS-CVR)
Adult
Two Mode
High
(PTS-CVR)
Adult
Two Mode
(PTS-CVR)
Pediatric
Two Mode
0-100mmHg
(PTS-CVR)
Neonatal
Two Mode
(PTSCVR)
C.A.S.S.
Intermittent
Suction Unit
(PTS-ISU)
Pediatric
Intermittent
Suction Unit
(PTS-ISU)
Neonatal
Intermittent
Suction Unit
(PTS-ISU)
Adult
Page 4
Page 4
Page 6
Page 6
Page 8
Page 8
Page 8
Page 11
Page 11
Page 10
On (Regulate)
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Off
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Max (Full line)
Yes
Yes
No
No
No
No
No
No
No
No
Features
Mode Selection:
Intermittent
No
No
No
No
No
No
No
Yes
Yes
Yes
Continuous
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Regulated
Vacuum
Range
0-760 mmHg
Full Vacuum
0-760 mmHg
Full Vacuum
0-760 mmHg
Full Vacuum
Gauge Range
Analog
0-200 mmHg
Full Vacuum
Gauge Range
Digital
0-200 mmHg
Full Vacuum
0-760 mmHg
Vacuum
Overall
Dimensions
Standards
0-200 mmHg
Full Vacuum
0-100 mmHg
Vacuum
0-160 mmHg
Vacuum
+/- 1% of Full Scale
Low,Med,
High & Full
Vacuum
Flow rate
Weight
0-160 mmHg
Vacuum
0-100 mmHg
Vacuum
0-200 mmHg
Full Vacuum
+/- 5% Full Scale
Gauge
Accuracy
(Digital)
*MR
Conditional
0-760 mmHg
Vacuum
0-160 mmHg
Vacuum
Gauge
Accuracy
(Analog)
Gauge
Color Band
Specifications are subject to change. Please see
“Instructions for Use” manual for the most current regulated vacuum
ranges.
High
Low,Med,
High & Full
Vacuum
High
0-80 LPM without fittings at full increase
Yes
Yes
Yes
+/- 1% of Full Scale
0-59 LPM without fittings at full increase
Yes
Low,Med,
High & Full
Vacuum
Low,Med,High
No
No
No
Intermittent - 0-8 LPM
Continuous - 0-80 LPM
without fittings at full increase
No
No
15 oz.
20 oz.
5.7”H x 3.0”W x 4.8”D
6.5”H x 2.8”W x 4.8”D
Yes
Conforms to ISO 10079-3 and ASTM F960 for all except Two Mode Low Continuous
Parts Repair
Kits
Available
Service
Manual
Available
* Effective January 2005, units noted above as MR Conditional conform to the guidelines recommended by the FDA and ASTM in a 3T environment
(outside MR scanner bore). Use in higher Tesla fields is not recommended. Fittings and adapters fitted to this product may have mild magnetic attraction.
To ensure MR compatibility, only adapters and fittings tested and designated by Ohio Medical for the configurations listed as MR Conditional should be
used. Any substitution or changes must be evaluated in accordance with your hospital policy.
Digital Regulators are NOT MR Conditional.
All products may not be available in all regulatory jurisdictions.
Information/products in this catalog subject to change without notice.
3