Download Vacuum Regulators
Transcript
Push-To-SetTM Vacuum Regulators The Ohio Medical® line of Push-To-SetTM (PTS) vacuum regulators are specially designed for a variety of clinical applications including airway, nasogastric, surgical suctioning and C.A.S.S. procedures as well as chest drainage. Ohio Medical has been providing suction products to healthcare facilities worldwide for 50 years. Our vacuum regulators continue to be the preferred choice for maximum performance, high quality and minimum maintenance. The vacuum regulators below are used in product configurations on pages 4-13. All Push-To-SetTM Vacuum Regulators feature: • Patented Push-To-SetTM technology for accuracy and • Safe, moderate to high flow rates safety • Glow-in-the-dark gauge needle and increments (Analog) • Quick-To-Max - 1 to 2 turns to full wall vacuum for • Dependable, time-tested gauges and controls emergency resuscitation • Low maintenance, easy service • Lightweight, compact design • Comprehensive ten year parts and labor warranty (Regulator only) • Impact-resistant, corrosion-free, rugged construction • Now available with a Digital Gauge (+/- 1% Gauge Accuracy) • Easy-to-read, color-coded gauge ranges • “Battery Low” Indicator on the digital gauge Three Mode Continuous (PTS-CVR) Adult Three Mode High (PTS-CVR) Adult Two Mode Continuous (PTS-CVR) Adult Two Mode High (PTS-CVR) Adult Two Mode (PTS-CVR) Pediatric Two Mode 0-100mmHg (PTS-CVR) Neonatal Two Mode (PTSCVR) C.A.S.S. Intermittent Suction Unit (PTS-ISU) Pediatric Intermittent Suction Unit (PTS-ISU) Neonatal Intermittent Suction Unit (PTS-ISU) Adult Page 4 Page 4 Page 6 Page 6 Page 8 Page 8 Page 8 Page 11 Page 11 Page 10 On (Regulate) Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Off Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Max (Full line) Yes Yes No No No No No No No No Features Mode Selection: Intermittent No No No No No No No Yes Yes Yes Continuous Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Regulated Vacuum Range 0-760 mmHg Full Vacuum 0-760 mmHg Full Vacuum 0-760 mmHg Full Vacuum Gauge Range Analog 0-200 mmHg Full Vacuum Gauge Range Digital 0-200 mmHg Full Vacuum 0-760 mmHg Vacuum Overall Dimensions Standards 0-200 mmHg Full Vacuum 0-100 mmHg Vacuum 0-160 mmHg Vacuum +/- 1% of Full Scale Low,Med, High & Full Vacuum Flow rate Weight 0-160 mmHg Vacuum 0-100 mmHg Vacuum 0-200 mmHg Full Vacuum +/- 5% Full Scale Gauge Accuracy (Digital) *MR Conditional 0-760 mmHg Vacuum 0-160 mmHg Vacuum Gauge Accuracy (Analog) Gauge Color Band Specifications are subject to change. Please see “Instructions for Use” manual for the most current regulated vacuum ranges. High Low,Med, High & Full Vacuum High 0-80 LPM without fittings at full increase Yes Yes Yes +/- 1% of Full Scale 0-59 LPM without fittings at full increase Yes Low,Med, High & Full Vacuum Low,Med,High No No No Intermittent - 0-8 LPM Continuous - 0-80 LPM without fittings at full increase No No 15 oz. 20 oz. 5.7”H x 3.0”W x 4.8”D 6.5”H x 2.8”W x 4.8”D Yes Conforms to ISO 10079-3 and ASTM F960 for all except Two Mode Low Continuous Parts Repair Kits Available Service Manual Available * Effective January 2005, units noted above as MR Conditional conform to the guidelines recommended by the FDA and ASTM in a 3T environment (outside MR scanner bore). Use in higher Tesla fields is not recommended. Fittings and adapters fitted to this product may have mild magnetic attraction. To ensure MR compatibility, only adapters and fittings tested and designated by Ohio Medical for the configurations listed as MR Conditional should be used. Any substitution or changes must be evaluated in accordance with your hospital policy. Digital Regulators are NOT MR Conditional. All products may not be available in all regulatory jurisdictions. Information/products in this catalog subject to change without notice. 3