Issue: Patient Temperature is not displayed on the screen
•
Ensure that the connection between the temperature probe and the cable is secure.
•
Ensure the probe and associated cable is connected to Patient Temperature 1 and not Patient
Temperature 2 Connector.
•
Ensure that the cable connection to the back of the Arctic Sun Control Module is secure.
•
If the temperature cable is wet, replace with a new cable.
•
Note: If temperature is out of range three dashes (---) will appear on the screen.
Issue: Low Temperature Alert/Alarm
•
If the patient temperature is 31°C (87.8°F) and water temperature is lower than patient temperature and
decreasing, an alarm condition will occur and revert to Stop Mode. If this occurs, follow the suggestions
in Appendix B and C.
•
From the Stop Menu, press the up or down arrows to the Advanced Settings screen. Press Enter. Using
the up and down arrows, scroll to the Low Patient Temperature Alert. If necessary, adjust the temperature
setting for the particular patient or procedure.
Issue: High Temperature Alert/Alarm
•
If the patient temperature is 39.5°C (103.1°F) and water temperature is higher than patient temperature and
increasing, an alarm condition will occur and revert to Stop Mode. If this occurs, follow the suggestions in
Appendix B and C on Alarms/Alerts.
•
From the Stop Menu, press the up or down arrows to the Advanced Settings Key. Press Enter. Using the
up and down arrows, scroll to the High Patient Temperature Alert. If necessary, adjust the temperature
setting for the particular patient or procedure.