Download Section 1 - Introduction

Transcript
5173574-2-100 rev 2
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging
System Conformance Statement for DICOM V3.0
X–Ray Filtration Module
Table 3-24: X–Ray Filtration Module Attributes
Attribute Name
Contour filter type
Tag
Type
(0018,1160)
3
Attribute Description
Type of filter(s) inserted into the X-Ray beam (e.g.
wedges). Defined Terms: STRIP WEDGE
BUTTERFLY MULTIPLE NONE
XA Positioner Module
Table 3-25: XA Positioner Module Attributes
Attribute Name
Tag
Type
Distance Source to Detector
(0018,1110)
3
Internally generated by the acquisition system.
Distance Source to Patient
(0018,1111)
3
Internally generated by the acquisition system.
Estimated Radiographic Magnification
Factor
(0018,1114)
3
Calculated from (0018,1110) and (0018,1111)
Positioner Motion
(0018,1500)
2C
Positioner Primary Angle
(0018,1510)
2
For multi–frame images, value of the first frame.
Positioner Secondary Angle
(0018,1511)
2
For multi–frame images, value of the first frame.
Positioner Primary Angle Increment
(0018,1520)
2C
Value of the RAO/LAO increments relative to the
first frame. If positioner motions is STATIC, will be
sent zero length.
Positioner Secondary Angle Increment
(0018,1521)
2C
Value of the CRA/CAU increments relative to the
first frame. If positioner motions is STATIC, will be
sent zero length.
Angle value 1
(0019,XX01)
3
Positioner angle for L arm (in degrees). Left
moving is positive.
Angle value 2
(0019,XX02)
3
Positioner angle for Pivot arm (in degrees). Left
moving is positive.
Angle value 3
(0019,XX03)
3
Positioner angle for C arm (in degrees). Head
moving is positive.
Angle 1 increment
(0019,XX97)
3
Incremental change in L arm with respect to the
angle of the first frame. Present only if positioner
motion is DYNAMIC.
Angle 2 increment
(0019,XX98)
3
Incremental change in Pivot arm with respect to
the angle of the first frame. Present only if
positioner motion is DYNAMIC.
Angle 3 increment
(0019,XX99)
3
Incremental change in C arm with respect to the
angle of the first frame. Present only if positioner
motion is DYNAMIC.
47/80
Attribute Description
Will be DYNAMIC if Pivot or C_ARM moves. If no
motion, or if L only moves, will be sent as STATIC.