Using the Options Menu Viewing Current Pump Personality Feature Set Note: When Flow Check is selected from the Options Menu, the status is displayed on the Main Display for 10 seconds (Figure 4-4). Note: If Flow Check is enabled in the Configuration Utility, the Flow Check display appears on the Main Display screen whenever the pump is running. Note: To check whether the back pressure from a downstream occlusion has decreased, view the Flow Check display when the pump is stopped. If all five triangles are filled, the downstream occlusion condition remains. Figure 4-4 Flow Check Main Display Viewing Current Pump Personality Feature Set From the Options Menu, use the ר keys to highlight the name of the current Personality feature set and press the Select soft key (Figure 4-5). The resulting screens allow you to view all the configuration settings for the current Personality feature set. See “Advanced Features,” 5-1 for additional information on selecting a Personality feature set. If the current Personality feature set shown is Permanent Settings, custom configurations are not selected. See “Configurable Options,” 9-1 for factory settings. 4 Figure 4-5 Highlight Name of Current Personality Feature Set Printed from the Baxter Intranet. THE USER IS RESPONSIBLE FOR CHECKING THE CURRENT ISSUE DATE/CHANGE NUMBER BEFORE USING THIS DOCUMENT. 07-19-D4-764 COLLEAGUE VOLUMETRIC INFUSION PUMP OPERATOR’S MANUAL 4-3